April 2021
Cause
The wireless radio has detected a failure or stopped communicating.
Recommended actions
Reset the device.
Probe malfunction
Alert
Probe malfunction
Cause
Measurement values are not consistent with the selected probe type.
Recommended actions
Instrument board measurement failure
Alert
Instrument board measurement failure
Cause
Due to an internal error, a measurement was not performed correctly.
Recommended actions
Reset the device.
Instrument board software error
Alert
Instrument board software error
Cause
Unable to read the software version on the instrument board.
Recommended actions
Quick Start Guide
1. Check that the correct probe type is set and perform a
measurement.
2. Check the probe wiring and connections and perform a
measurement.
3. Reset the device.
4. Inspect the probe and probe cable for defects and replace it, if
necessary.
1. Check that the firmware version on the instrument board is
correct.
Quick Start Guide
35