Preparations For Testing; Entering The Device Data; Checking The Devices Self Testing - Weinmann Homecare MEDUCORE Easy Service And Repair Instructions

Automatic external defibrillator. device nos. 1000 to 1340 from device no. 1341 with ilcor 2005
Table of Contents

Advertisement

4.2 Preparations for testing

EasyView
EasyView-Service
EasyView-Light

4.3 Entering the device data

4.4 Checking the devices self testing

10
Checking the device
1. Boot up PC.
2. Start the PC software.
Overview of software versions:
EasyView
Reading, editing and saving usage data,
modifying device settings.
EasyView-Service
Works like EasyView, facility for programming the
serial number.
EasyView-Light
Works like EasyView; device settings cannot be
modified.
Enter the serial number of the device (Device-No.) and
the manufacturing date into the test records.
1. Insert rechargeable or Battery-Pack.
2. Open the device lid.
Request:
– The device is giving off a short double peeping
tone.
– All LEDs light up briefly.
– The device announces clearly and completely
the spoken order "This device will assist you.".
– The green status LED lights up.
3. Close the device lid.