Performing the test
Energy
162 J
WAV> VFIB PLAY HDR esc

4.7 Checking the procedure display and capacitor discharge

12
Checking the device
4. Press the arrow button under ENERG.
The indicator for the Energy Modus appears in the
display.
5. To simulate venticular fibrillation: Press the arrow
button under WAV until VFIB appears in the
display.
1. Open the lid of the device.
2. Plug the defibrillator tester electrode plug into the
device.
Requirement:
– The device detects ventricular fibrillation and
after a maximum of 10 sec clearly gives the
spoken message "Shock required".
3. The device prepares to give the shock.
Requirement:
– The current consumption of the device is be-
tween 6 A and 12 A.
4. Note the current consumption in the test report.
5. After the spoken message: "Press the flashing
shock key", activate this.
Requirement:
– The shock is triggered by activation of the
shock button.
– The shock is given. The device clearly gives the
spoken message "Shock was delivered".
– The energy discharged is between
168 J ± 10 % (Low Energy). The value is
shown in the display of the defibrillator tester.
Note:
Only the testing of the "Low Energy" shock is necessary.
If it is also required to test the "High Energy", the energy
must be 298 J ± 10 %.
In this case, ventricular fibrillation must be set on the
defibrillator tester as described above. Return to the main
menu via esc.
Note:
Do not press the shock key during this test!
1. Simulate ventricular fibrillation with the defibrilla-
tor-tester (Signal Vfib).