Advertisement

4. Checking the device

4.1 Testing parts needed

Important!
The device must undergo the following tests after every
repair and every maintenance according to WM 40006,
and the results must be entered into the test records.
Additionally the test can be used for fault finding in the
device.
If you detect errors or deviations from defined values
in the course of your final testing you must not redeploy
MEDUCORE Easy before these errors have been
eliminated.
You can determine possible causes of such errors and
recommended countermeasures by referring to
chapter "7. Malfunctions and Rectification" on
page 22.
We recommend that you generally keep the following
ready for use:
Emergency-Set WM 15460
Battery-Pack WM 40155
Rechargeable Battery-Pack WM 40150
Electrodes, packed WM 40116
Important!
The electrodes have a limited shelf life. Please make a
note of the expiry date on the packaging.
Fully charged Rechargeable battery Pack or
Battery-Pack
Defibrillator-Tester (see also 10.4, page 38)
IR-Adapter WM 22498
PC with installed software EasyView WM 40192
Multimeter, Measuring range up to 20 A
Battery adapter WM 40008
Test line MEDUCORE Easy WM 40454