Datex-Ohmeda AS/3 CM and CS/3 CMC
77.
S-ANE99(A)
L-ARK99(A)
S-ICU99(A)
?
Network Board, B-NET/ B-CMNET/ B-UPINET
78.
NOTE: During battery operation the battery symbol overrides the network symbol.
NOTE: If the network symbol does not appear, check the status of the network.
?
72
Document No. 896 623
Display Trends
Press the
Check that there is enough trend information available for the monitored parameters.
Erase the trends:
Reset Case
- Reset ALL - YES
Check that the trends have been erased by pressing the
Reload trends from the Data card by pressing the
Prev. Case, pressing the ComWheel on the last saved file (the file information is shown
at the bottom of the menu) and selecting Load.
Wait until the message 'Loading from Mem. Module' disappears then check that trend
information is available again by pressing the
Admit/ Discharge
- Discharge - YES
Check that the trends have been erased by pressing the
selecting Trends.
Reload trends from the Data card by pressing the
Admit Patient - Patient from Card, pressing the ComWheel on the last saved file (the
file information is shown at the bottom of the menu) and selecting Load.
Wait until the message 'Loading from Mem. Module' disappears then check that trend
information is available again by pressing the
Check that the Mon-Net cable connector and the Identification plug are clean and intact,
then connect them to the Network Board.
Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbol
appears under the clock on the upper right-hand corner of the screen. Also a message
regarding the connected Datex-Ohmeda Central should appear in the message field on the
screen.
Pt Data & Trends
membrane key/
Patient Data
Display Trends
Admit/ Discharge
Pt Data & Trends
key and select Trends.
Display Trends
key again.
key, selecting Load
key
.
Pt Data & Trends
key and
key, selecting
key
.