3.
4.
5.
CAUTION
The Memory Board, B-CMMEM comprises sensitive integrated circuits that can be
damaged by an electrostatic discharge. Careful handling of the board is therefore
essential.
Unscrew the two nuts holding the side plates in position and remove the plates. Put the
memory card collar in the place of the side plates.
Unscrew the screws holding the bottom plate in position and remove the plate.
Remove the Memory Board, B-CMMEM, including the connection board and the memory
board, from the protective anti static packaging. Always hold the boards by the edges and
wear a wrist grounding strap.
General Service Guide
2
27
Document No. 896 623