Beds require an effective maintenance program, we recommend checking these items annually . Use this sheet for
your records . Keep on file .
CHECKLIST
_____ All fasteners secure (reference all assembly prints) .
_____ Engage brake pedal and push on the bed to ensure all casters lock securely .
_____ "Brake Not Set" LED on the footboard and head end siderails when brakes are not engaged .
_____ Locking steer caster engages and disengages properly (2131 Model only) .
_____ Siderails move, latch and stow properly .
_____ CPR release working properly .
_____ I . V . pole working properly (if on bed) .
_____ No cracks or splits in head or footboards .
_____ No rips or cracks in mattress cover .
_____ All functions on head end siderails working properly (including LED's) .
_____ All functions on footboard working properly (including LED's) .
_____ Scale and bed exit system working properly .
_____ Night light working properly .
_____ Power cord not frayed .
_____ No cables worn or pinched .
_____ All electrical connections tight .
_____ All grounds secure to the frame .
_____ Ground impedance not more than 100 milliohms .
_____ Current leakage not more than 300 microamps .
_____ Engage drive wheel and ensure it is operating properly (ZOOM® option - 2141 model only) .
_____ Motion release switches working properly (ZOOM® option - 2141 model only) .
_____ Confirm head end ZOOM® handle functionality (2141 model only) .
_____ Confirm battery powered functionality .
_____ Ensure ground chains are clean, intact, and have at least two links touching the floor .
_____ Check fowler angle for accuracy 0
_____ Check gatch angle for accuracy 0
_____ Check foot section angle for accuracy 0
_____ Siderail switches working properly (iBED Awareness option) .
_____ iBED Awareness Light Bar LED's on footboard and siderails working properly (iBED Awareness option) .
_____ Inspect footboard control labeling for signs of degradation .
_____ Ensure calibration of the touch screen is accurate .
_____ Ensure calibration of the bed is accurate .
Bed Serial Number:
Completed by: ___________________________________________
www.stryker.com
Preventative Maintenance
- 70
.
0
0
0
- 15
0
.
0
- 50
0
.
________________________
________________________
________________________
________________________
2131-009-005 REV A
________________________
________________________
________________________
________________________
Date: _________________
Return To Table of Contents
1-79
Need help?
Do you have a question about the InTouch FL27 and is the answer not in the manual?