OPTIONAL ZOOM
®
DRIVE SYSTEM (MODEL 2141 ONLY)
In addition to the previous warnings and cautions, all of the following warnings and cautions apply to units equipped
with the ZOOM
®
.
WARNING
• The 2141 InTouch™ Bed is intended for use by trained hospital personnel only . Failure to properly train personnel
could result in injury .
• USE CAUTION while maneuvering the unit with the drive wheel activated . Always ensure there are no obstacles
near the unit while the drive wheel is activated . Injury to the patient, user or bystanders or damage to the frame or
surrounding equipment could occur if the unit collides with an obstacle .
• Use caution when transporting the unit down halls, through doors, in and out of elevators, etc . Damage to the
siderails or other parts of the unit could occur if the unit comes in contact with walls or door frames .
• Put the drive wheel in the neutral position and release the brake before pushing the unit manually . For 2141 models;
push the Brake Off button to disengage drive wheel (ZOOM
push the unit manually with the drive wheel engaged . The unit will be difficult to push and injury could result .
• If unanticipated motion occurs, unplug the power cord from the wall socket, push the battery power on/off switch
to the "OFF" (0) position (the LED will not be illuminated), actuate the drive wheel pedal to the neutral position
and call maintenance .
• The power save mode is activated after one hour on battery power with no motion release switch activation .
Functions including Bed Exit, Scale and Motion will cease to operate when the unit enters the power save mode .
Injury to the patient could occur if proper patient monitoring protocol is not observed .
• Always unplug the power cord and push the battery power on/off switch to the "OFF" (0) position before service
or cleaning . When working under the frame, always support the litter frame to prevent injury in case the Bed Down
switch is accidently activated .
• Battery posts, terminals and related accessories contain lead and lead compounds, chemicals known to the State
of California to cause cancer and birth defects or other reproductive harm . Wash hands after handling .
• Do not modify the InTouch™ Model 2141 bed . Modifying the unit can cause unpredictable operation resulting in
injury to the patient or user . Modifying the unit will also void this warranty .
Return To Table of Contents
1-10
Summary of Safety Precautions
2131-009-005 REV A
) before pushing the unit manually . Do not attempt to
®
www.stryker.com