Page 1
GoBed ® MedSurg Bed Model FL28C Operations Manual For Parts or Technical Assistance: USA: 1-800-327-0770 (option 2) Canada: 1-888-233-6888 2010/06 72-0908E OM FL28C REV C www.stryker.com...
Refer to your local distributor for return and/or collection systems available in your country. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Note This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
PRODUCT DESCRIPTION The GoBed® II MedSurg Bed, Model FL28C offers a high level of security, intuitive operation and enhanced mobility that allows you to support varying types of patients populations with ease and efficiency. • Safe. Full siderail coverage, low bed height and the Chaperone center-of-gravity bed exit system combine to provide our patients with the highest level of protection.
700 hPa 500 hPa Stryker reserves the right to change specifications without notice. • Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations. • Operating environment recommended to ensure the scale system (optional) precision.
• Emergency crank handle • Oxygen bottle holder • Bed extender and pad (with and without scale option) • Removable I.V. poles • Fixed I.V. poles • Traction sleeves and adapters Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Before operating the GoBed II MedSurg Bed, Model FL28C, it is important to read and understand all information in this manual. Carefully read and strictly follow the safety guidelines listed below. It is important that all users have been trained and educated on the inherent hazards associated with the usage of electric beds.
Page 12
• Because individual beds may have different options, foot boards should not be moved from one bed to another. Mixing foot boards could result in unpredictable bed operation. • When servicing use only identical replacement parts provided by Stryker. • Make sure that the ground chain is in place, intact and it’s touching the floor (see Figure 1.9, page • Do not put anything under the bed.
II MedSurg Bed, Model FL28C is working properly before it is put into service. The following list will help ensure that each part of the bed is checked. • Perform a full visual inspection of the bed. If there are any damages, please contact your local Stryker Medical representative.
To engage the wheel brakes, fully depress the side of the pedal identified by the red arrow labeled “BRAKE”. To disengage the wheel brakes, toggle the pedal to the neutral position. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
The bed may be equipped with an optional photoelectric night light to illuminate the floor area around the bed. The night light turns on as the room lights dim. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Use only hospital grade electric equipment consuming 5A or less with the auxiliary power outlet (optional). The use of standard electric equipment may bring the current leakage to an unacceptable level for hospital equipment. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
(see D1, D2 and D3, “Foot Board Control Panel Guide” page 19) before removing the foot board. WARNING Do not put anything under the bed. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Figure 2.13B A: Press to raise Gatch C: Press to raise Fowler B: Press to lower Gatch D: Press to lower Fowler G: Press to activate Nurse Call (optional) Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Press to obtain the desired reverse Trendelenburg (head up, foot down) position. To return the sleep surface to the horizontal position, simply use the bed Hi-Lo controls to either raise or lower the sleep surface to its limits. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
• Should the message “No Scale Found” appear on the screen during the normal course of operation of the scale, press the enter key. Data will reappear on the screen and the current operation will resume. If the message remains, contact Technical Support. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Gain or Loss XXX.X lb G/L XX.X To access the patient’s weight variation anytime, go to the Gain or Loss menu through the Menu Up/Down controls once the scale is activated. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 22
• Select the weight unit using the Menu Up or Down control and press ENTER. The system will return to the Scale mode with the weight displayed in the unit chosen. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
• If the bed is not equipped with the scale system, remove the patient from the bed; change the equipment and zero the system (see ¨Zeroing the System¨ on page 24). The Bed Exit system may be activated once the patient returns to the bed. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
• Press and hold the ARM/DISARM key (K) until the larger zone LED (left LED) begins to flash, then release the control (do not touch the bed while the LED flashes). When the zone LED turns off, the system is zeroed. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 25
• If the bed is not equipped with the scale system, remove the patient from the bed; change the equipment and zero the system (see ¨Zeroing the System¨ on page 24). The Bed Exit system may be activated once the patient returns to the bed. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 26
Notes Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Fixed Two Stage I.V. Pole, Foldable FA64171 See page 44 Fixed Three Stage I.V. Pole, Foldable FA64170 See page 46 FA64188 FA64189 Traction Adapters See page 48 FA64190 FA64191 Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Insert the cable into the wire holder (B) to prevent the cable from pulling out of the receptacle. Attach the pendant control to the bed sheets using the clip. OPERATION Speaker/Microphone for Nurse Call On/Off switch and Volume adjustment Nurse Call button TV On/Off button Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Attach the pendant control to the bed sheets using the clip. OPERATION Speaker/Microphone for Nurse Call On/Off switch and Volume adjustment Nurse Call button Room Light On/Off button Reading Light On/Off button TV On/Off button Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Hook the pendant control on to the siderail. OPERATION Raise head section Lower head section Raise bed Lower bed Raise foot section Lower foot section Lock ring Cable connector Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Tighten the connector lock ring. Hook the pendant control on to the siderail. OPERATION Raise head section Lower head section Raise foot section Lower foot section Lock ring Cable connector Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
To place the tray into its stored position, pull the lock pin (A), lower the tray to its lowest position and release the pin. Note: Secure the monitor to the tray using the Velcro strap. WARNING The maximum load capacity of the tray is 40 lb (18 kg). Lock pin Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
To adjust the trapeze (H) height, shorten or lengthen the strap (J). If you must detach the strap to free the trapeze, ensure the strap is properly reinserted in the fastener. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 34
WARNING The overhead trapeze should not be used when in the storage position. This may result in injury to the patient or dam- age to the equipment. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Insert the crank into the appropriate aperture located at the foot end of the bed and follow the icon indications to adjust as desired. Remove the cranks after use and store it at the head end of the bed. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
When using a half bed length tent type, ensure the siderails are outside the oxygen tent and the oxygen tent should not extend below the mattress support level. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Unplug the bed from the wall socket. Remove the foot board. Configuration with 80” Mattress Configuration with 84” Mattress Note the position of the foot end mattress retainer. Support assembled Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 38
Insert the foot board into the mounting sockets on the bed extender. Verify proper operation of the unit before returning it to service. WARNING The bed extender must be removed before using the emergency crank (optional). Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Before installing the extender, adjustments must be made depending on the length of mattress being used: Configuration with 80” Mattress Configuration with 84” Mattress Note the position of the foot end mattress retainer. Support assembled Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 40
Insert the foot board into the mounting sockets on the bed extender. Verify proper operation of the unit before returning it to service. WARNING The bed extender must be removed before using the emergency crank (optional). Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
• Tighten the locking sleeve. Hang the I.V. bags. Tightening Sleeve WARNING The weight of the I.V. bags must not exceed 11 lb (5 kg). Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
• Tighten the locking sleeve. Hang the I.V. bags. WARNING The weight of the I.V. bags must not exceed 11 lb (5 kg) per hook. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Rotate the I.V. pole hangers (D) to the desired position and hang the I.V. bag(s). WARNING The weight of the I.V. bags must not exceed 40 lb (18 kg). Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
To store the I.V. pole, turn the latch (G) clockwise until the telescoping section (E) lowers. Lift up on the pole to remove it from the receptacle. Fold the pole down and store it on the bracket. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 45
Insert the rollers (G) on each bracket and secure using the new hexagonal bolts (I) and nuts (K). Note: When tightening the nuts, make sure that there is enough room for the rollers to spin freely. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Turn the latch (K) clockwise to lower the first section. Lift the pole to remove it from the receptacle. Fold the pole down and store it on the bracket. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 47
Insert the rollers (G) on each bracket and secure using the new hexagonal bolts (I) and nuts (K). Note: When tightening the nuts, make sure that there is enough room for the rollers to spin freely. Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Note: The bolt is coated with “Scotch Grip”. The coating becomes less effective once the bolt is tightened and removed thereafter. Replace the bolt if the traction adapter is removed and reinstalled on another bed. Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
_____ Current leakage not more than 100 µA (microamps). _____ Make sure that the ground chain is in place, intact and it’s touching the floor (see Figure 1.9, page Bed Serial Number: Completed by: _______________________________________ Date: _________________ Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products suggested above are used to clean Stryker patient care equipment, measures must be taken to insure the bed is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the bed will leave a corrosive residue on the surface of the bed, possibly causing premature corrosion of critical components.
Recycling Passport OL280005 (Reference Only) Item Recycling/Material Code Important Information QDF28-0265 Circuit Board QDF5095 Buzzer Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 52
Recycling Passport OL280120 (Reference Only) Item Recycling/Material Code Important Information 28-0769L Motor 28-0770 Motor 25-0768 Motor QDF23-0315) Circuit Board Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 53
Recycling Passport OL280121 (Reference Only) Item Recycling/Material Code Important Information 28-0769L Motor 28-0770 Motor 25-0768 Motor QP23-0315 Circuit Board QDF14-1160) Transformer Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 54
Recycling Passport OL250029 (Reference Only) Item Recycling/Material Code Important Information QDF9025 Circuit Breaker Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 55
Recycling Passport OL280097, OL280098, OL280100 (Reference Only) Item Recycling/Material Code Important Information QDF20-0126 Circuit Board Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Page 56
Recycling Passport OL280035, OL280101, OL280102 (Reference Only) Item Recycling/Material Code Important Information QDF20-0121 Circuit Board Return To Table of Contents 72-0908E OM FL28C REV B www.stryker.com...
Page 57
Recycling Passport OP280001, OP280002, OP280003 (Reference Only) Item Recycling/Material Code Important Information QDF28-0266 Circuit Board Return To Table of Contents www.stryker.com 72-0908E OM FL28C REV C...
Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by Stryker, products or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product materially and...
Stryker authorized parts used Service during regular business hours (8−5) * Does not include maintenance due to abuse or for any disposable items. Stryker reserves the right to change options without notice. Stryker Medical also offers personalized service contracts. Pricing is determined by age, location, model and condition of product.
Guidance and Manufacturer’s declaration - Electromagnetic Immunity ® The GoBed II MedSurg Bed, Model FL28C is suitable for use in the electromagnetic environment specified below. ® The customer or the user of the GoBed II MedSurg Bed, Model FL28C should assure that it is used in such an environment.
Page 61
II MedSurg Bed, Model FL28C. ® The GoBed II MedSurg Bed, Model FL28C is intended for use in an electromagnetic environment in which ® radiated RF disturbances are controlled. The customer or the user of the GoBed II MedSurg Bed, Model FL28C can help prevent electromagnetic interferences by maintaining a minimum distance between portable ®...
Page 62
The GoBed ® II MedSurg Bed, Model FL28C is suited for use in the electromagnetic environment specified below. The customer or the user of the GoBed II MedSurg Bed, Model FL28C should assure that it is used in such an ®...
Page 63
Guidance and Manufacturer’s declaration - Electromagnetic Emissions ® The GoBed II MedSurg Bed, Model FL28C is intended for use in an electromagnetic environment specified ® below. The customer or the user of the GoBed II MedSurg Bed, Model FL28C should assure that it is used in such an environment.
Page 65
Stryker France Stryker France ZAC Satolas Green Pusignan ZAC Satolas Green Pusignan Av. De Satolas Green Av. De Satolas Green 69881 MEYZIEU Cedex 69881 MEYZIEU Cedex France France 2010/06 2007/XX 72-0908E OM FL28C REV C XXXX-XXX-XXX REV X www.stryker.com www.stryker.com...