Stryker InTouch FL27 Operation Manual

Critical care bed
Hide thumbs Also See for InTouch FL27:
Table of Contents

Advertisement

InTouch™ Critical
Care Bed
Model FL27 (2131/2141)
Version 2.3
Operations Manual
For parts or technical assistance call:
USA: 1-800-327-0770 (option 2)
Canada: 1-888-233-6888
2009/06
2131-209-001 REV A
www.stryker.com

Advertisement

Table of Contents
loading

Summary of Contents for Stryker InTouch FL27

  • Page 1 InTouch™ Critical Care Bed Model FL27 (2131/2141) Version 2.3 Operations Manual For parts or technical assistance call: USA: 1-800-327-0770 (option 2) Canada: 1-888-233-6888 2009/06 2131-209-001 REV A www.stryker.com...
  • Page 3: Table Of Contents

    Main Menu: Bed Controls ............. . www.stryker.com...
  • Page 4 Right Fit Oxygen Bottle Holder - FA64203........... 2131-209-001 REV A www.stryker.com...
  • Page 5 Recycling Passports ..............www.stryker.com...
  • Page 6: Introduction

    INTENDED USE This manual is designed to assist you with the operation of Stryker InTouch™ Critical Care Bed Model FL27 (2131/2141). Carefully read this manual thoroughly before using the equipment or beginning maintenance on it. To ensure safe operation of this equipment, it is recommended that methods and procedures be established for educating and training staff on the safe operation of this bed.
  • Page 7: Specifications

    • Fowler 15 seconds • Foot 60 seconds • Seat 30 seconds Drive System Speed (Optional ZOOM ® Drive) • Forward 3.6 mph 5.8 km/h • Backwards 2.5 mph 4.0 km/h Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 8: Specifications (Continued)

    3.0 V 220mAh Lithium Battery, Size 20mm - Varta Int. CR2032 Coin Cell PC Battery Holder, Size 20mm - MDP Int. BA2032 Stryker reserves the right to change specifications without notice. Specifications listed are approximate and may vary slightly from unit to unit or by power supply fluctuations.
  • Page 9: Symbols

    Refer to your local distributor for return and/or collection systems available in your country. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 10: Summary Of Safety Precautions

    Unplug the bed power cord from the wall when oxygen administering equipment is used. When using a half bed-length tent type, ensure that the siderails are outside the oxygen tent and that the tent does not extend below the mattress support level. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 11 • Because individual beds may have different options, footboards should not be moved from one bed to another. Mixing footboards could result in unpredictable bed operation. • When servicing use only identical replacement parts provided by Stryker. Note • Throughout this Operations Manual, the words “right” and “left” refer to the right and left sides of a patient lying on his/her back on the bed.
  • Page 12: Optional Zoom ® Drive System (Model 2141 Only)

    • Do not modify the InTouch™ Model 2141 bed. Modifying the unit can cause unpredictable operation resulting in injury to the patient or user. Modifying the unit will also void this warranty. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 13: Setup Procedures

    Fowler fully up using the “Fowler Up” control. • Verify the following optional equipment for proper operation: 120V auxiliary outlet, night light, emergency crank operation, etc. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 14: Optional Xprt Therapy Mattress

    Install the power cord in the two (2) retaining clips. 10. Fasten the three (3) retaining straps. 11. Carefully rotate the foot end control box and the mattress into the flat position. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 15: Applying The Linens

    ZOOM® drive (2141 Model only) should now be engaged. Depress down fully on the center of the pedal until the letter N in BLUE is visible in the clear window. The ZOOM drive should now be disengaged. ® Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 16: Modes Of Operation

    Mode if the batteries no longer have power and need to be recharged. The system then goes to the Initialization Mode which goes back to the Normal Mode. Note The time needed to charge both batteries completely is approximately 8 hours. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 17: Base Operation Guide

    To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the Drive mode activated; the Drive wheel cannot swivel. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 18: Litter Operation Guide

    Improperly adjusted restraint straps can cause serious injury to a patient. It is the responsibility of the attending medical personnel to determine proper use of the restraint straps and restraint strap locations. Stryker is not responsible for the type and/or use of restraint straps on any of Stryker’s products.
  • Page 19: Nurse Call Usage

    If the footboard must be removed and the bed electric functions remain accessible through the siderails, ensure that the siderail controls are operational by deactivating any lockout activated before removing the footboard. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 20: Positioning The Siderails

    To Raise the Siderails (Figure 1.1) • Pull on the siderail until it is locked in the fully raised position; a “click” will be heard to indicate it is locked. (See figure below) Figure 1.1 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 21: Bed Positioning

    These positions can be achieved by pressing their respective buttons on the “Nurse Control”. (See page 22). OPTIONAL BED EXTENDER The bed can be expanded using a Bed Extender. For further information see the Maintenance Manual. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 22: Siderail Operation Guide

    Press on this button to raise the height of the bed to the desired level. Bed Height Down Press on this button to lower the height of the bed to the desired level. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 23: Bed Positioning Descriptions

    • A minimum clearance of 20 inches is required to allow the Trendelenburg and Reverse Trendelenburg positions to be executed. • These positions cannot be executed when one or more locks is activated. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 24: Siderail Brake/Motion Control Panel, Nurse Control

    Press this button to start a treatment with the bed in the HOB HOB 30 Degree Position 30 Degree position. Note The brake can also be set manually. See page 17 for more information. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 25: Optional Pendant Control

    Press this button to raise the Gatch section of the bed to the desired angle. Gatch Down Press this button to lower the Gatch section of the bed to the desired angle. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 26: Footboard Operation Guide

    Brake Activation/Indicator button is illuminated when the system is activated. Brake Off Press this button to activate the Brake Off system. This button Activation/Indicator is illuminated when the system is activated. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 27: System Message Indicator Lights

    If a submenu is opened and no further action is taken by the user, the LCD will shut off completely after 4 minutes (4 minutes = 2 minutes at 100% intensity + 2 minutes at 10% intensity). Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 28: Main Menu: Patient Information

    • Unit/Room : pressing on the pencil tool icon in the bottom left of the Patient Information screen allows the operator to enter the unit and room number that the patient is in. Figure 3.0 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 29: Patient Information, New Patient

    BMI, displaying the patient’s ID and displaying the last Braden scale score. Calculate BMI Display Last Display Braden Scale Patient ID Score Cancel Figure 3.2 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 30: Main Menu: Bed Controls

    This indicator shows the actual height of the bed. Note • When pressed the buttons turn grey. • The bed height is actually the distance between the floor and the bed support mattress. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 31: Main Menu: Scale

    Press this button to select the measure wanted (pounds or Lbs/Kg kilograms). Press on this button to start the Delayed Weight. This will give Delayed Weight the user the time to add or remove accessories. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 32 The user can change the delayed weight measuring value to Kg or Lbs by pressing on the “Lb/Kg” button located at the bottom of the Delayed Weight screen as shown in the Scale menu. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 33: Zero Button

    The “Zero” function will reset all the values for the different patient’s weights. All the data stored in the touch screen concerning the weight of the patient will be lost. Close Figure 5.5 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 34: Weight History Button

    The system can keep a maximum of fifty (50) entries in memory. Any new value added after the entry will have the first value deleted following Previous Next Cancel the “First In, First Out” rule. Figure 5.7 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 35: Change Equipment Button

    Negative values cannot be obtained when using the touch screen menus. Figure 5.10 Message Message Add/remove equipment Change equipment success Please press Enter when done Enter Close Close Figure 5.11.1 Figure 5.11 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 36: Change Weight Button

    0.1 lb or kg. If the arrows are 0.0 lbs pressed longer than just a touch, the incrementation/ decrementation will change by 1 lb or kg. To save the new weight, press “Ok”. Cancel Figure 5.14 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 37: Gain Or Loss Button

    When the “Do not touch bed” message disappears, the “Gain or Loss reset success!” message will be displayed as shown in Figure 5.17 Gain Or Loss reset success! Close Figure 5.17 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 38 Notes Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 39: Main Menu: Bed Exit

    • The blue color on one of the zones indicates that this is the zone which is actually activated. • When the zone color turns red, it means that there is a Bed Exit Alarm. In that case the Event Manager window will appear. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 40 When the volume or tone is changed from its current setting, Test Volume / Tone you may press the test button to hear the new tone setting or the new volume adjustment. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 41: Main Menu: Support Surfaces

    If one of the siderails is unlocked during a rotation therapy (ex.: rotation and vibration is being executed simultaneously), both therapies will stop instantly. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 42: Positionpro™ Mattress Specifications

    Matress Turn In Progress the PositionPRO™ uses the fowler and Trendelenburg tilt sensors of the InTouch™ to sense the angle of the Back fowler. Unit/Rm : Oct. 25,2007 12:00 AM Figure 7.5 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 43 This shows the percentage of power left in the battery. Battery Each battery block indicates 25 percent of the power. Backlight Press this button to access the Backlight menu. History Press this button to access the History menu. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 44: Time Setup Button

    Language Language Button English When the “Language” button is pressed on the Options screen, Figure 8.2 will be displayed. You Español can now select the desired language. Français Cancel Figure 8.2 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 45: Advanced Options Button

    Voice option is activated, sound will be played through the speakers when the ZOOM® is activated. If the Awareness and/or Reminder options are selected, the alarms will be sent through the bed Nurse Call signal. Figure 8.3 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 46: History Button

    40 ~ 49° 12:00 24:00 50 ~ 59° 12:00 24:00 60 ~ 69° 12:00 24:00 70 ~ 79° 12:00 24:00 80 ~ 89° 12:00 24:00 HOB 30 Reset Close History Figure 8.7 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 47 Height History needs to be reset, press the “Reset” 00:00 00:00 button. 00:00 00:00 00:00 00:00 00:00 00:00 00:00 00:00 Low height Reset Close History Figure 8.9 Nurse Control Backlight Button Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 48: Nurse Control Backlight Button

    Close Clear log Trend 11.6 5411 5172 838363 Back Brake unable to elec. Unit/Rm : Oct. 25,2007 12:00 AM 12:00 AM Unit/Rm : Oct. 25,2007 12:00 AM Figure 8.13 Figure 8.12 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 49: Main Menu: Lockouts

    • If the bed is in Trendelenburg or Reverse Trendelenburg the bed will be locked in this position. • When one of the locks is enabled, the indicator on the footboard will turn on so the user is aware that some motion may not be activated. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 50: Main Menu: Ibed

    60 seconds without correcting the conditions, it goes automatically back to the “Smart Bed Position” screen until the corrections are made. Unit/Rm : Oct. 25,2007 Oct. 25,2007 12:00 AM Figure 10.2 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 51: Ibed, Protocol Reminder

    The default screen is the one-time reminder, but the Hours One time task user can change it to “Repeated Task” by pressing on the “Repeated Task” button. Nov 13, 2008 Back Unit/Rm : Oct. 25,2007 12:00 AM Figure 10.5 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 52: Ibed, Protocol Reminder (Continued)

    2008/11/13 - 10:29 AM When the View Reminder’s Log button is pressed on the Reminder screen, Figure 10.8 will be displayed. Clear Back Unit/Rm : Oct. 25,2007 12:00 AM Figure 10.8 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 53: Ibed, Head Of Bed (Hob) History

    24 hours. If the Bed 00:00 00:00 00:00 00:00 Height History needs to be reset, press the “Reset” 00:00 00:00 button. 00:00 00:00 00:00 00:00 Low height Reset Close History Figure 10.10b Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 54: Ibed, Calculator

    Figure 10.13 will be displayed. When the operator completes their conversions, they may switch back to the “Calculator” by pressing on the “Calculator” button. Unit/Rm #: Oct. 25,2007 12:00 AM Figure 10.13 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 55: Ibed, Translations

    To increase the audio, press the “+” button. To decrease the audio, press the “-” button. Note: If the iAudio option is not equipped on your bed, then the audio controls buttons will not be dispalyed in Figure 10.14 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 56: Ibed, Documentation

    Press the View button and a log will be displayed for that date. You may view specific logged information by pressing either the Bed Exit, Reminder, Scale, iBed or System buttons. Figure 10.15 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 57: Ibed, Braden Scale

    Braden Scale Result Summary 1. Sensory Perception 2. Moisture 3. Activity 4. Mobility 5. Nutrition 6. Friction & Shear Total Score : Close Unit/Rm : Oct. 25,2007 12:00 AM Figure 10.22c Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 58: Ibed, Sound Therapy

    Note: The “Music Note” icon will continue to be displayed on all menu screens if Sound Therapy Back is playing. Unit/Rm : Oct. 25,2007 12:00 AM Figure 10.23b Music Note Icon Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 59: Zoom ® Drive Operations (Model 2141)

    • If the bed battery charge level become too low, the motorized wheel will retract (raise) while driving, or won’t go down. Figure 26 Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 60: Optional Bed Accessories

    Traction Sleeve 8” x 3/4” FA64218 See page 77 Traction Sleeve 6-1/2” x 3/4” FA64219 See page 77 Wall Saver Cable FA64208 See page 78 X-Ray Cassette Holder FA64205 See page 79 Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 61: Bed Extender - Fa64234

    • Ensure connector ‘A’ is securely fixed (see Figure B). • Do not sit on the bed extender. This may cause the bed to flip. Section 2 Connector A Section 1 Figure B Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 62: Permanent I.v. Pole - Fa64221

    Note: The bolt used is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 63 To store the I.V. pole, turn the latch (C) clockwise until section (A) lowers. Lift up on the pole to remove it from its receptacle, fold it and store it on the bracket. CAUTION The weight of the I.V. bags should not exceed 40 lbs (18 kg). Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 64: Dual Head End Permanent I.v. Pole - Fa64202

    Note: The bolt used is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 65 To store the I.V. pole, turn the latch (C) clockwise until section (A) lowers. Lift up on the pole to remove it from its receptacle, fold it and store it on the bracket. CAUTION The weight of the I.V. bags should not exceed 40 lbs (18 kg). Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 66: Line Management System (Accessory Clip) - Fa64210

    • Do not use with tubing of a diameter greater than 0.75” (19 mm). • Sterilize accessory after each use (see procedure in Maintenance Manual) • Make sure that the clip is stable when installing. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 67 Notes Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 68: Monitor Tray - Fa64214

    Figure A Figure C - bracket mounted (back) Figure D - bracket mounted (front) Figure B - foot end Figure E - secure bracket Figure F - install tray with screw knob Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 69 Figure L - secure with straps WARNING • The maximum load capacity of the tray is 40 lbs (18kg) • The maximum load capacity of the tubular support is 150 lbs (68 kg) Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 70: Upright Oxygen Bottle Holder - Fa64187

    Secure the Support Bar to the head end of the bed by inserting the security chain through the Support Bar hole located at the end of the Support Bar. Support Bar Support Bar Hole Security Chain Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 71: Right Fit Oxygen Bottle Holder - Fa64203

    Note: The cover is reversible. It can be placed so that the opening is on the right or on the left of the bed to ease the use. Insert the oxygen bottle here. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 72: Pendant With Motion Control Only - Fa64209

    Plug the cable connector into connector (A) located on the right of the upper frame. Install the removable pendant on the right foot siderail. Insert the removable pendant in in this space. Slide it until it clips. Note: Illustration may differ according to bed model. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 73: Pendant With Motion Control/Nurse Call - Fa64194

    Plug the cable connector into connector (A) located on the right of the upper frame. Install the removable pendant on the right foot siderail. Insert the removable pendant in in this space. Slide it until it clips. Note: Illustration may differ according to bed model. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 74: Pendant With Motion Control/Smart Tv - Fa64195

    Plug the cable connector into connector (A) located on the right of the upper frame. Install the removable pendant on the right foot siderail. Insert the removable pendant in in this space. Slide it until it clips. Note: Illustration may differ according to bed model. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 75: Pendant With Motion Control/Nc/Smart Tv - Fa64193

    Plug the cable connector into connector (A) located on the right of the upper frame. Install the removable pendant on the right foot siderail. Insert the removable pendant in in this space. Slide it until it clips. Note: Illustration may differ according to bed model. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 76: Pendant Clip - Fa64186

    The Optional Pendant Clip was designed to support the removable pendant (FA64193, FA64194, FA64195, FA64209). INSTALLATION Required Tools: None Raise the clip prong (A) and install the clip on the bed at a safe and useful location. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 77: Traction Sleeve - Fa64215 Through Fa64219

    Note: The bolt used at step 2 is coated with “Scotch Grip”. This type of bolt must be replaced by a new identical bolt if it is removed following a first installation. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 78: Wall Saver - Fa64208

    Screw the connector’s screws at the bed location. Join both cables. Insert the connector in the other one on the cable from the hospital. Screw the connector’s screws to the hospital connector. Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 79: X-Ray Cassette Holder - Fa64205

    Note: Make sure to close the X-ray cassette holder support after using it to avoid damaging it when there is a movement of the fowler section or of the list system. Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 80: Cleaning

    SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY. If the products described above are used to clean Stryker patient care equipment, measures must be taken to ensure the beds are wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning.
  • Page 81: Preventative Maintenance

    _____ Ensure calibration of the touch screen is accurate. _____ Ensure calibration of the bed is accurate. Bed Serial Number: ________________________ ________________________ ________________________ ________________________ ________________________ ________________________ ________________________ ________________________ Completed by: ___________________________________________ Date: _________________ Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 82: Warranty

    Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by Stryker, products or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product materially and...
  • Page 83: Service Contract Programs

    Stryker authorized parts used Service during regular business hours (8−5) * Does not include maintenance due to abuse or for any disposable items. Stryker reserves the right to change options without notice. Stryker Medical also offers personalized service contracts. Pricing is determined by age, location, model and condition of product.
  • Page 84: Recycling Passports

    Recycling Passports Assembly Part number: 27-2287/27-2288 (Reference Only) (27-2287 shown) Item Recycling/Material Code Important Information (QDF27-1097) Brake Control Board (QDF27-1099) Siderail Nurse Control Board Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 85 Recycling Passports Assembly Part number: OL270236 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1429) Nurse Call Board (QDF75-0280) Can-Conn Board Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 86 Recycling Passports Assembly Part number: OL270032 (Reference Only) Item Recycling/Material Code Important Information (QDF2060) Smart TV Board Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 87 Recycling Passports Assembly Part number: OL270060 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1562) iBed Electronic Board Lens Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 88 Recycling Passports Assembly Part number: OL270181 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1099) Siderail Nurse Control Board Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 89 Recycling Passports Assembly Part number: 27-2547-XXX (Reference Only) Item Recycling/Material Code Important Information 27-2548 (ZOOM Control Board) QDF9136 (ZOOM Load Cell) Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 90 Recycling Passports Assembly Part number: OL270167 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1562) iBed Electronic Board Lens Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 91 Recycling Passports Assembly Part number: OL270079 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0010) Menu Board (QDF27-1097) Brake Control Board Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 92 Recycling Passports Assembly Part number: L27-019 (Reference Only) Item Recycling/Material Code Important Information (27-2477) Fowler Angle Sensor Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 93 Recycling Passports Assembly Part number: L27-019 (Reference Only) Item Recycling/Material Code Important Information (27-2477) Gatch Angle Sensor Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 94 Recycling Passports Assembly Part number: 27-1200 (Reference Only) Item Recycling/Material Code Important Information (27-2477) Foot Angle Sensor Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 95 Recycling Passports Assembly Part number: L27-043 (Reference Only) Item Recycling/Material Code Important Information (27-2477) Base Angle Sensor Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 96 Recycling Passports Assembly Part number: L27-021 (Reference Only) Item Recycling/Material Code Important Information (27-2477) Trend Angle Sensor Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 97 Recycling Passports Assembly Part Number: 27-2478 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0110) DC Power Control Board (QDF9188) 12V 18Ah Battery Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 98 Recycling Passports Assembly Part number: 27-2546 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1430) ZOOM Board (QDF75-0240) ZOOM Interface Board Return To Table of Contents 2131-209-001 REV A www.stryker.com...
  • Page 99 Recycling Passports Assembly Part number: L27-014 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1372) Litter Load Cell Return To Table of Contents www.stryker.com 2131-209-001 REV A...
  • Page 101 UNITED STATES Stryker Medical 3800 E. Centre Ave., Portage, Michigan USA 49002 CANADA Stryker Canada 45 Innovation drive Hamilton, Ontario Canada L9H 7L8 European Representative Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex France 2009/06 2131-209-001 REV A www.stryker.com...