InTouch Critical Care Bed ® FL27 (2131/2141) Version 4.0 with Wi-Fi / Isolibrium™ (2971) support surface Maintenance Manual 2018/03 H.0 2141-209-002 REV H www.stryker.com...
Refer to your local distributor for return and/or collection systems available in your country. Non-ionizing radiation; i.e. RF transmitter (Wi-Fi) Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Symbols iBed Locator is connected iBed Locator is not connected Wireless Network is connected Wireless Network is not connected Support surface call maintenance Serial number Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Note This provides special information to make maintenance easier or important instructions clearer. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Introduction This manual assists you with the operation or maintenance of the Stryker Model FL27 (2131/2141) InTouch ® Critical Care bed. Read this manual thoroughly before operating or maintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenance of this product.
Stryker beds. The desired bed parameters will be set by clinicians at the bedside. The iBed Wireless ® software is intended to be used only with specifically enabled Stryker beds that have been verified and validated with the iBed ®...
Speed to level bed from any position • Fowler 15 seconds • Foot and Seat 60 seconds Zoom ® Speed motorized • Forward 2.98 mph 4.8 km/h drive (Model • Backwards 1.79 mph 2.88 km/h 2141) (option) Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Isolibrium™) (35” x 6” x 4”) • With Bed Extender DM64197 (XPRT™) (35” x 6” x 5.5”) 35” x 90” x 5.5” 88,9 cm x 228,6 cm x 13,9 cm Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Scale Accuracy 64 °F (18 °C) Specifications listed are approximate and may vary slightly from product to product or by power supply fluctuations. Stryker reserves the right to change specifications without notice. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Roller bumpers Footboard control panel Siderail release levers Headboard Siderails Head end control panel (optional) Support surface (optional) InTouch touch screen Zoom drive system (model 2141) (optional) Isolated foley bag hooks Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Portage, Michigan 49002 To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication. SERIAL NUMBER LOCATION You can find the serial number plate behind the patient right siderail near the foot end of the product.
Do not apply the electronic brake to slow or stop the product while it is in motion. • Always keep the product in a low, horizontal position with the siderails fully raised and locked when transporting a patient. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Always determine the proper use of the restraint straps and restraint strap locations. Improperly adjusted restraint straps can cause serious injury to a patient. Stryker is not responsible for the type or use of restraint straps on any of Stryker’s products.
Fluids can cause unpredictable operation and decreased functionality of any electrical product. Do not return the product to service until it is completely dry and has been thoroughly tested for safe operation. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Always immediately replace batteries that have corrosion at the terminals, display cracking, have expanded or bulging sides, or no longer can maintain a full charge. • Always use only Stryker authorized batteries when replacing the batteries. Use of non-Stryker batteries may lead to unpredictable system performance. •...
All electronic service parts will be shipped in static shielding bags. Do not open the bags until you have completed steps 2 and 3 of the Static Protection Procedure. Note: Always ship back circuit boards to Stryker in the same antistatic bags that the new boards were originally shipped in.
Always unplug the product power cord from the wall when using oxygen administering equipment. Possible fire hazard exists when this product is used with oxygen administering equipment other than nasal, mask type, or half bed-length tent type. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
You must also remap the iBed Locator if you change the room or location information after initial installation. If you have any problems during the iBed Wireless installation process, contact Stryker Technical Support at 1–800–327−0770.
Release the Main Menu button and you will be taken into the Configuration Screen shown in Figure 2. Brake Main Menu Drive Figure 1: Footboard Control Panel Figure 2: Configuration Screen Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Press the Bed Calibration button on the Configuration Screen (refer to Figure 3). After pressing the Bed Calibration button, Figure 4 will be displayed. Press “NEXT” to continue with calibration. Figure 4: Bed Calibration - WARNING Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Bed Height Down (Head Lift Down) and Reverse Trendelenburg (Foot Lift Down) buttons simultaneously as shown in Figure 7. Bed Height Down (Head Lift Down) QDF27-2237 Figure 7: HE Siderail Control Panel - (Right Side Shown) Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Figure 8 When step 1 of the calibration procedure has completed, step 2 of the calibration procedure will begin and Figure 9 will be displayed. Figure 9: Bed Calibration - Step 2 Return To Table of Contents www.stryker.com 2141-209-002 REV H...
12. Press the Next button when done. The “Do Do Not Not Touch Bed” screen will appear as shown Touch Bed in Figure 12. Cancel Figure 12: Do Not Touch Bed Screen Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Page 29
Trendelenburg (Foot Lift Up) button as shown in Figure 14. Verify +12 degrees +/- 0.1 with the inclinometer you previously placed on the litter frame in step 11. QDF27-2237 Figure 14: Head Siderail Control Panel Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Page 30
17. When step 3 of the calibration procedure has completed, step 4 of the calibration procedure will begin and Figure 16 will be displayed as shown below. Figure 16: Bed Calibration - Step 4 Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Page 31
20. Press the Next button when done. Bed Calibration Do Not 21. The “Do Not Touch Bed” screen will appear as Touch Bed shown in Figure 18. Cancel Figure 18: Do Not Touch Bed Screen Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Page 32
Trendelenburg (Foot Lift Up) button and Bed Height Up (Head Lift Up) button as shown in Figure 20. QDF27-2237 Figure 20: Head Siderail Control Panel - (Right Side Shown) Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Page 33
25. Press the Next button when done. Bed Calibration 26. The “Do Not Touch Bed” screen will appear as Do Not shown in Figure 22. Touch Bed Cancel Figure 22: Do Not Touch Bed Screen Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Page 34
Foot Down button until limit is met (see Figure 24) . CAUTION: The Gatch must be raised to the highest height prior to running the Foot Down or damage could occur. QDF27-2237 Figure 24: Calibration Complete Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Page 35
16”, verify the display readings are still at zero degrees. Note: If readings are not all zero, you will need to recalibrate the bed by repeating steps 1-30. Figure 25: Calibration Complete Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Nurse SRR - Trend Hilo Foot - Forward Footboard - Hob Nurse Head - Leg up Foot - Forward Close Close Figure 27d: Motor Figure 27e: Pressed Buttons Figure 27f: Signal Values Return To Table of Contents 2141-209-002 REV H www.stryker.com...
NOTE: The TOUCH SCREEN CALIBRATION SCREEN is a validation of the touch screen’s calibration. When the user presses anywhere on the screen, the coordinates x and y are displayed on the lower part of the screen. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Turn the battery switch to the ON (|) position. Note: After you cycle power, InTouch reboots to the Patient Information screen. Figure 30: Network Parameters Figure 31: User Authentication Figure 33: Advanced Figure 32: Security Return To Table of Contents 2141-209-002 REV H www.stryker.com...
OFF (O) then unplug the power cord from the wall outlet. Plug the power cord back into the wall outlet then turn the battery disconnect switch back ON (|). Test bed functionality prior to returning the bed into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
ON (|). Test bed functionality prior to returning the bed into service. Figure 38: Cycle Power Screen Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Therapy History (C) (press and hold for a minimum of 5 seconds) Figure 40: Diagnostic Entry Combination If you have entered into the diagnostic menu and need to exit, tap X (Figure 41). Figure 41: Diagnostic Main Screen Sample Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Always tap Clear Error History after troubleshooting is complete • If you clear the error history while there is an active error, the active error does not show in the Error History tab until the active error is fixed. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
To enter the Advanced Signal Values screen, tap Advanced Signal Values on the Diagnostics screen (Figure 44). Figure 44: Diagnostic Main Screen - Sample To exit the Advanced Signal Values screen, tap Back to return to the Diagnostics screen (Figure 45). Figure 45: Advanced Signal Values Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Turn Left • Turn Right • Low Air Loss To run diagnostics, tap start on the lower portion of the Diagnostics screen (Figure 46). Figure 46: Diagnostic Main Screen - Sample Return To Table of Contents 2141-209-002 REV H www.stryker.com...
When the test is complete, the results of each test section is indicated by a red cross when there is a failure or a green check mark when successful (Figure 48). Note: When a failure is first detected, all following pressure tests will be marked as a failure. Figure 48: Diagnostic Test Result - Sample Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Appropriate disinfectant for nylon Velcro should be determined by hospital protocol. CLEANING A SUPPORT SURFACE To clean and disinfect a support surface, see the cleaning and disinfecting instructions in the support surface operations manual. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Preventive Maintenance At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more frequently based on your level of product usage. Remove product from service before performing preventive maintenance. Preventive maintenance should only be performed by trained or certified personnel.
Labels for legibility, proper adherence, and integrity _____ iBed Wireless Module and IR Module are intact and footboard icons are displaying (iBed Wireless option) Serial Number: Completed by: _______________________________________ Date: _________________ Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Quick Reference Replacement Parts These parts are currently available for purchase. Call Stryker Customer Service: 1–800–327–0770 for availability and pricing. Part Name Part Number Electrical Components Actuator, Brake QDF27-1227 Actuator, Foot QDF27-1216 Actuator, Fowler QDF27-1214 Actuator, Gatch QDF27-1215 Actuator, Lift, Foot End...
Caster, Steer (2131 Model Only - both foot end casters) RD27-1971 Headboard Assembly 27-2583K Footboard Assembly without iBed 27-2285K Footboard Assembly with iBed and iAudio 27-2824K Footboard Assembly with iBed and Wi-Fi 27-2827K Footboard Assembly with iBed, without iAudio 27-2826K Return To Table of Contents 2141-209-002 REV H www.stryker.com...
III. If (see CV Table) VAC is present, replace the transformer assembly. A. Check fuse F1 on the CPU/Power board. If fuse is good, replace CPU/Power board. If fuse does not have continuity, replace the fuse (littelfuse 215008.P). Return To Table of Contents www.stryker.com 2141-209-002 REV H...
J4 while pressing the trend button. Red lead to pin 1, black lead to pin 2. a. If voltage is present, replace the motor. b. If voltage is not present, replace the CPU/Power board. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
J3 while pressing the gatch up button. Red lead to pin 1, black lead to pin 2. a. If voltage is present, replace the motor. b. If voltage is not present, replace the CPU/Power board. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
J1 while pressing the foot down button. Black lead to pin 1, red lead to pin 2. a. If voltage is present, replace the motor. b. If voltage is not present, replace the CPU/Power board. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
25). a. If recalibration does not work, replace the fowler angle sensor and recalibrate. b. If replacement of the fowler angle sensor did not resolve the problem, replace the CPU/Power board. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
If the brake motor does not run down when the brake button is pushed, check for 24VDC at connector J7 while pressing the brake button. Black lead to pin 1, red lead to pin 2. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
, Figure 51 appears. When the scale system is unstable due to a problem with scale electronics or excessive patient movement, the weight reading is orange (Figure 52). Figure 51 Figure 52 Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Cmd WO safe from pat pend H Optional pendant control has a network communication error (Check network connections). Cmd WO safe from TS Touch screen has a network communication error (Check network connections). Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Network B1 has a wiring issue (Check cabling). Safe WO Cmd from B2 Network B2 has a wiring issue (Check cabling). Scale ADC Error Scale chip not calibrated. DC Control Board needs to be replaced. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Sensor Board not calibrated 0200 GUI CANopen Heartbeat Lost 0400 EEPROM corruption 0500 Ambient temperature sensor failure 0501 Ambient temperature out of range 0600 Pressure sensor failure 0601 Seat Left pressure sensor failure Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Back Left inflate valve current when OFF 1406 Back Left deflate valve current when OFF 1407 Back Right inflate valve current when OFF 1408 Back Right deflate valve current when OFF Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Turn Right deflate timeout 1700 Reference voltage hardware failure 1800 Weight input out of range 1900 HOB angle out of range for Lateral Rotation 1901 HOB angle out of range for Turn Assist Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Pin 2 - White • Drive Bed Unplugged Zoom® Drive Actuator (Model 2141 Only) 25VDC Pin 2 - White Pin 1 - Black 25VDC Pin 1 - Black Pin 2 - White Return To Table of Contents 2141-209-002 REV H www.stryker.com...
HI-LO HEAD M- HEAD HI-LO MOTOR EXIT HI-LO HEAD M+ GATCH M- GATCH M+ GATCH MOTOR EXIT FOOT M- FOOT MOTOR EXIT FOOT M+ WARNING SIGNAL - EXIT WARNING SIGNAL WARNING SIGNAL + Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Bed Electrical Diagram V IN - V IN + ZOOM ACTIVE ZOOM ACTIVE GND Enable Enable BATTERY - BATTERIES BATTERY + Return To Table of Contents 2141-209-002 REV H www.stryker.com...
5 (Manual) 4 (Neutral) SIGNAL NEUTRAL 2 (Brake Pos) 1 (+5V) Analog SIGNAL "BRAKE POSITION" SIGNAL MANUAL ENGAGE 3 .NC RIGHT FOOT SIDERAIL NC GNDD 1 .NO RIGHT FOOT SIDERAIL NO Return To Table of Contents 2141-209-002 REV H www.stryker.com...
NURSE ANSWER LIGHT PRIORITY NO/NC PRIORITY COM OPTION 3A COM TV COM (TV -) SPEAKER LOW COM AUDIO SHIELD RADIO NO/NC Pillow Speaker Nurse Call and Speaker Left Side Pillow Speaker Return To Table of Contents www.stryker.com 2141-209-002 REV H...
IIBED LOW HGT ALERTING TV COM (TV -) SPEAKER LOW COM AUDIO SHIELD AWARENESS COMMON IR - TX IR-RX FEED BACK IR CONNECTOR IR - TX IR CONNECTOR IR-RX FEED BACK Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Serial Interface Connector Cable QDF27-2212 Head Siderail Cable QDF2060 Smart TV Board QDF21-2901 SIP-N-PUFF Cable QDF27-1099 Siderail & Head Nurse Control Board 3 QDF75-0010 Menu Touch Screen Board QDF9573 XPRT™ Mattress Plug Return To Table of Contents 2141-209-002 REV H www.stryker.com...
10. Unplug the cable quick connect and remove the actuator. 11. Reverse the steps to install the new actuator. 12. Test all bed functionality prior to putting bed back in service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Unplug the cable quick connect and remove the actuator. 10. Reverse the steps to install the new actuator. 11. Test all bed functionality prior to putting bed back in service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
1.00 ±0.06 pliers to reattach the spring to the center of the clevis pin. 17. Test all bed functionality prior to putting bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Note: If new actuator shaft does not line up with mounting holes, mount the base of the actuator then plug the actuator in and run it electrically in or out until lined up. 10. Test all bed functionality prior to putting bed back in service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Note: If new motor shaft does not line up with mounting holes, mount the base of the actuator then plug the actuator in and run it electrically in or out until lined up. 12. Test all bed functionality prior to putting bed back in service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Using a pry bar or a 2 x 4, pry upward on the Zoom® drive actuator frame and push the actuator clevis pins out. Remove actuator. Reverse steps to install new actuator. Test all bed functionality prior to putting bed back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Number Configuration procedure on page 40). 15. Recalibrate the bed by referring to the Bed Calibration procedure on page 16. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Ensure cables do not get pinched when securing the head end frame cover together. 11. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 12. Test all bed functionality prior to putting bed back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Do not pinch the cable when installing the load cell. 15. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 16. Test all bed functionality prior to putting bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
15. Reverse the procedure to install the new display. 16. Recalibrate the touch screen, by referencing the Touch Screen Calibration procedure on page 17. Test all bed functionality before returning the bed to service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
10. Disconnect the brake control cable from the brake control board. 11. Remove the brake control board. 12. Reverse the procedure to install the new brake control board. 13. Test all bed functionality before returning the bed to service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Unplug the cable going to the function selection/LED board. 10. Reverse the procedures to install the new function selection/LED board. 11. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
17. Reverse the procedure to install the new touch screen. 18. Recalibrate the touch screen, by referencing the Touch Screen Calibration procedure on page 19. Test all bed functionality before returning the bed to service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Remove the battery and replace with new battery (see figure below). Reverse the above steps to reassemble. 10. Referencing the operation manual complete the Time / Date Setup procedure. Battery Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Note: Make sure to install new wire ties in the same location. Recalibrate the bed (refer to Bed Calibration procedures on page 25). Test all bed functionality prior to putting the bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Note: New battery information should be facing the right side when installing and upright when installed. 11. Test all bed functionality while plugged into wall outlet and when unplugged prior to putting back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Note: Do not over-tighten the two screws. 10. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 11. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Note: Do not over tighten the two screws. 10. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 11. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Note: Do not over-tighten the two screws. 10. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 11. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Note: Do not over-tighten the two screws. 10. Recalibrate the bed (refer to Bed Calibration procedures on page 25). 11. Test all bed functionality prior to putting the bed back into service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Reverse the steps to install new trend angle sensor. Note: Do not over-tighten the two screws. Recalibrate the bed (refer to Bed Calibration procedures on page 25). Test all bed functionality prior to putting the bed back into service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
15. Reverse steps 12 through 1 to install the new headwall communication board. 16. Test all bed functionality before returning the bed to service. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Note: Route both right and left IR module cables into the head end frame through the matching right and left Zoom handle cable cutouts in the head end frame. ® 16. Test all bed functionality before returning the bed to service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
10. Using a Phillips screwdriver, remove the two screws that secure the Wi-Fi board to the LCD interface board. 11. Unlock the Wi-Fi board from the LCD interface board by pressing in until it clicks. 12. Remove the Wi-Fi board from the LCD interface board. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Press the Wi-Fi icon in the bottom left corner of the touch screen and verify the bed received a proper hospital issued IP address. 19. Test all bed functionality before returning the bed to service. Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Base Assembly, Hood L27-026 Rev J (Reference Only) 16 NOTE 1 HEAD HAND TIGHTENED, THE SCREW HEAD NOTE 1: MUST BE WELL SEATED ON THE COVER. FOOT Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Foot Base Cover QP27-1093 Center Base Cover QP27-1105 Head Base Cover QP27-1461 Stopper in Base VV23A1G24HL Tapping Screw VV83A9G16 Tapping Screw VVB4A1O24 Thread Rolling Bolt VV84A1I16 Flanged Hex Head Tapping Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Base Assembly, Brake The angle sensor orientation is very important. Wire should exit towards the bottom. DETAIL B Tyrap positioning DETAIL A View of Assembled Brake System Return To Table of Contents 2141-209-002 REV H www.stryker.com...
To assemble casters, direct the smile (chrome part) toward the left side of the bed. DETAIL A Item Part No. Part Name Qty. RD27-2787 5.5” Caster Brake/Neutral/Neutral RD27-2788 5.5” Caster Brake/Neutral/Drive Return To Table of Contents 2141-209-002 REV H www.stryker.com...
RIGHT SEE DETAIL A NEUTRAL To assemble casters, direct smile (chromed part) toward the left side of the bed. DETAIL A Item Part No. Part Name Qty. RD27-2787 5.5” Caster Brake/Neutral/Neutral Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Angle Sensor Assembly 27-2477 Rev 0 (Reference Only) Item Part No. Part Name Qty. QDF75-0140 Angle Sensor QP27-1210 Angle Sensor Case QP27-1211 Bottom Lid for Tilt Sensor Return To Table of Contents 2141-209-002 REV H www.stryker.com...
DC Control Board see drawing Litters Wiring for opening see photo #13 in litter 1 J131 1 J115 see photo #1 1 J126 see photo #11 see photo #6 Leave 2” Loop Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Bed must be in high position • Gatch must be at 15 degrees • Foot Litter must be at 45 degrees see photo #8 Leave 2” loop. see photo #5 see photo #7 and 12 Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Be sure to place the connectors one beside the other and in a at position. PHOTO #1 PHOTO #2 PHOTO #5 PHOTO #4 PHOTO #3 PHOTO #7 PHOTO #8 PHOTO #6 PHOTO #9 PHOTO #10 Return To Table of Contents 2141-209-002 REV H www.stryker.com...
QDF27-2182 Bornes J170A QDF9159 No Manual Switch QDF27-2182 Bornes J170B QDF9159 COM Manual Switch QDF27-2182 J146 QDF27-1252 Hi-lo Head Motor QDF27-2182 J147 QDF27-1251 Hi-lo Foot Motor QDF27-2182 J149 QDF27-1227 Brake Motor Return To Table of Contents www.stryker.com 2141-209-002 REV H...
J24 DC Control Board Base Assembly, Electrical - 27-2687 Rev B (Reference only) Item Part No. Part Name Qty. QDF27-2181 Wires Harness #2 QDF27-2182 Base Structure Extender QDF9518 Wire Tie QDF9523 Wire Tie Return To Table of Contents 2141-209-002 REV H www.stryker.com...
CABLE CONNECTION QDF27-1646 TABLE Cable No. Connector No. Cable No. Connector No. QDF27-1646 Mini Fit QDF27-1185 J132 QDF27-1646 Black Eyelets QDF9188 BAT #2 POLE - QDF27-1646 White Eyelets QDF9188 BAT #1 POLE + Return To Table of Contents www.stryker.com 2141-209-002 REV H...
® - OL270232 Rev C (Reference Only) Item Part No. Part Name Qty. 27-2547 Zoom® Handle Assembly 27-2546 Zoom® Drive Base Assembly QDF27-1646 Battery Wire QDF9518 Cable Attachment QDF27-1185 Wire Harness Return To Table of Contents 2141-209-002 REV H www.stryker.com...
27-2546 Rev-02 (Reference Only) PLACE THE HYDRAULIC GREASER TOWARD THE BOTTOM AND REMOVE THE HEX HEAD SCREW ON THE BEARING. 10 J3, J4, J5, J6, J10 30 J10 J150 HEAD FOOT Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Lift Assembly Head End INTERIOR OF THE SHOULDERS TO ASSEMBLE THIS PART THE ARROW DIRECTED TOWARDS THE HEAD OF THE BED LEFT RIGHT HEAD FOOT LEFT RIGHT Return To Table of Contents 2141-209-002 REV H www.stryker.com...
THE SPRING OF GREASE COMPLETELY INTERIOR OF THE SHOULDERS DETAIL A 39 33 THE ORIENTATION OF THIS SENSOR OF ANGLE IS IMPORTANT (THE WIRE MUST LEAVE DOWNWARDS) RIGHT LEFT HEAD FOOT LEFT RIGHT Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Stiffener Arm Steel Washer QP27-2019 Stiffener Arm Acetal Washer 27-2018Z Stiffener Arm Bearing 27-2053Z Clevis Pin Diameter QDF7878 Hitch Pin Diameter VW10C432002 Nylon Washer VE30A0G Nylon Locknut 27-2117 Spring in Base QDF2100 Round Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Litter Assembly, Interior Frame L27-055 Rev D (Reference Only) HEAD FOOT SECTION B-B GREASE PASS THE WIRE OVER THE RUBBER SLEEVE DETAIL A Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Spacer Head VW10A08 Flat Washer VB15A1N32 Bolt VE30A1N Nylon Hex Locknut M0019 Grease 27-2880 Foot end hardstop additional spacer 2 27-2881 Head end hardstop additional spacer 2 VV83A9G32 Pan head tapping screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Litter Assembly with Serial Port OL270278 Rev A (Reference Only) CONNECT ON J1 (TO QDF75-0270) Return To Table of Contents www.stryker.com 2141-209-002 REV H...
VR11H43 Pop Rivet VV83A9E12 Pan Head Tapping Screw VV83A9G12 Pan Head Tapping Screw VV83A9G16 Pan Head Tapping Screw VW10A12 Washer 27-2005P Hiding place cell VV33A0G24 Phillips Head Screw VE30A0G Nylon Nut Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Litter Assembly, Foot End 27-2688 Rev B (Reference Only) COVER ASSY 4.00 ±1.00 HEAD SEE ASSEMBLY OF THE COVER FOOT Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Litter Assembly, Foot End DIRECT WIRE TOWARDS THE INTERIOR OF THE BED DETAIL B THE ORIENTATION OF THIS ANGLE SENSOR IS IMPORTANT (THE WIRE MUST BE DOWNWARDS) Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Litter Assembly, Foot End ALWAYS ASSEMBLE THE TWO SPACES W/O LEG TOWARDS THE INTERIOR OF THE BED Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Litter Assembly, Standard Cabling L27-059 Rev B (Reference Only) Item Part No. Part Name Qty. 27-2687 Base and Lever Wiring 27-2695 Litter Wiring Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Cable No. Connector No. Cable No. Connector No. QDF27-1646 Mini Fit QDF75-0440 J12 DC Board QDF27-1646 Black Grommet QDF9188 Battery #2 Pole - QDF27-1646 White Grommet QDF9188 Battery #1 Pole + Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Left Head Cell QDF27-2213 QDF75-0440 J23 DC Board QDF27-2213 QDF75-0440 J26 DC Board QDF27-2213 J200A Black QDF9571 Line “L” QDF27-2213 J200B White QDF9571 Line “N” QDF27-2213 J200C Green/Yellow Ground Head Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Tapping Screw Flat Head Phillips VV83A9G16 Tapping Screw Pan Head Phillips 27-1723P Head Plate without Auxiliary Outlet QDF2155 Steel Cable Clamp VV33A0G16 Machine Screw Pan Head Phillips VV83A9G24 Pan Head Tapping Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Leave a 4” loop here. See Photo #6 See Photo #7 and #8 Leave a 4” loop here. See Photo #1, 2, 3 See Photo #4 See Photo #5 1.00 ±0.06 Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Shaft Support Bearing QP27-2065 CPR Pedal QRE23-0438 Siderail Spring VB15A1N44 Bolt VE30A0G Nylon Locknut VE30A1N Nylon Locknut VG10B0636 Spring Pin VV33A0G24 Pan Head Machine Screw VV83A9G12 Pan Head Tapping Screw VV83A9G16 Pan Head Return To Table of Contents www.stryker.com 2141-209-002 REV H...
VV83A9G12 Pan Head Tapping Screw QDF5096 Plastic Tie VE30A1O Nylon Hex Locknut VB15A1O32 Bolt 27-2790P Zoom Handle Bracket, Right ® 27-2789P Zoom ® Handle Bracket, Left QE71-1373 Coiled Power Cord Label Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Zoom ® Handle Assembly (Model 2141 Only) 27-2547-XXX Rev B (Reference Only) Note: Attach cables under the head box. Return To Table of Contents www.stryker.com 2141-209-002 REV H...
OL270077-XXX Rev D (Reference Only) TOP VIEW NOTE: XXX - indicates language choice (tri = English/French/Spanish; bil = English/French) Item Part No. Part Name Qty. QE71-0946-XXX Zoom Label ® QE71-1372 Zoom Handle Label ® Return To Table of Contents www.stryker.com 2141-209-002 REV H...
No Head End Control Panel Assembly - OL270019 Rev-01 (Reference Only) Item Part No. Part Name Qty. QDF27-1649 Head End Control Fascia without Motorized Zoom® Drive QP27-1650 Head Frame Plate VV23A9C12HL Pan Head Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
12” Network Cable QDF27-1099 Siderail Nurse Control Board QF27-2236-XXX Fascia, Head of Bed Control QDF9183 Electronic Board Button QP27-1111 Nurse Control Board Support VV23A9C12HL Pan Head Screw QE71-1340 Connection Isolation Sticker Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Siderail Mounting Assembly, Head End L27-039 Rev-01 (Reference Only) Head Foot Item Part No. Part Name Qty. VB18A1O32 Bolt VE30A1O Nylon Locknut 27-2287 Head End Siderail, Right 27-2288 Head End Siderail, Left Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Standard Siderail Assembly, Head End, Left 27-2288 Rev-04 (Reference Only) 29 T35+/-3.5 LUBRICATE T22+/-2 DETAIL A Return To Table of Contents 2141-209-002 REV H www.stryker.com...
NOTE: XXX - indicates language choice (tri = English/French/Spanish; bil = English/French) Item Part No. Part Name Qty. QDF27-0906-XXX Nurse Control Label, HE, Right QDF27-0907-XXX Nurse Control Label, HE, Left QDF27-2239-XXX Brake Control Label, HE, Right QDF27-2240-XXX Brake Control Label, HE, Left Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Room Interface Board QDF27-2175 Siderail Speaker, Left VV23A9C12HL Pan Head Tapping Screw VV83A9G16 Pan Head Tapping Screw QDF27-2211 Label, Left Communications QDF27-2196 Head Speaker QDF27-2216 Siderail Speaker, Right QDF27-2210 Label, Right Communications Return To Table of Contents www.stryker.com 2141-209-002 REV H...
QDF27-2198 Nurse Call and Speaker Fascia, Right QDF27-1429 Nurse Call Board QDF27-2199 Nurse Call and Speaker Fascia, Left QDF27-2216 Siderail Speaker, Right QDF27-2175 Siderail Speaker, Left QDF27-1682 Siderail Nurse Call Wires Return To Table of Contents www.stryker.com 2141-209-002 REV H...
IR Module, Right Label, iAudio/Comm., Left QDF27-2211 27-2661 IR Module, Left QDF27-2196 Head Speaker QDF75-0270 Room Interface Board QDF27-2210 Label, iAudio/Comm., Right VV83A9G36 Pan Head Phillips QDF27-2216 Siderail Speaker, Right QDF9518 Cable Tie Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Optional Siderail Assembly, Head End, with Speaker/with NC/ with iAudio/with iBed ® Wireless OL270302 Rev B (Reference Only) Return To Table of Contents www.stryker.com 2141-209-002 REV H...
TOP VIEW (HEAD FRAME COVER NOT SHOWN) NOTE 2-3-4 CABLE FROM RIGHT IR MODULE (ITEM 11) NOTE 2-3-4 TOP VIEW (HEAD FRAME COVER NOT SHOWN) LEFT IR MODULE CABLE (ITEM 13) Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Optional Siderail Assembly, Head End, with Speaker/with NC/ with iAudio/with iBed ® Wireless Path of IR Module Cables to the Room Interface Board NOTE 3 UNDER BED VIEW NOTE 3 NOTE 3 UNDER BED VIEW Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Left Siderail Speaker QDF27-1682 Siderail Nurse Call Wires 27-2661 IR Module, Left QDF75-0270 Room Interface Board 27-2662 IR Module, Right QDF9518 Cable Tie QP27-2735 IR Support Gasket, Left QP27-2734 IR Support Gasket, Right Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Siderail Mounting Assembly, Foot End L27-032 Rev-02 (Reference Only) Item Part No. Part Name Qty. 27-2289 Right Foot Siderail Assembly 27-2290 Left Foot Siderail Assembly VB18A1O32 Bolt VE30A1O Nylon Locknut Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Standard Siderail Assembly, Foot End, Right 27-2289 Rev-03 (Reference Only) PART QP28-0039 IS OF GRAY COLOR - 21 T35±3.5 CODE OF COLOR:10 TOWARDS FRAME 39 LUBRICATE T22 ±2 SEE SIDERAIL OPTION RACEWAY DETAIL A Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Black NC SWITCH Green LED1 LBS Orange +12V LED2 LBS Blue Note 1: Siderail without iBed option - See OL270059-XXX (page 200). Siderail with iBed option - See OL270060-XXX (page 201). Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Standard Siderail Assembly, Foot End, Left 27-2290 Rev-03 (Reference Only) PART QP28-0039 IS OF GRAY COLOR - CODE OF COLOR:10 18 T35±3.5 TOWARDS FRAME T22 ±2 SEE SIDERAIL OPTION RACEWAY DETAIL A Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Black NC SWITCH Green LED1 LBS Orange +12V LED2 LBS Blue Note 1: Siderail without iBed option - See OL270059-XXX (page 200). Siderail with iBed option - See OL270060-XXX (page 201). Return To Table of Contents 2141-209-002 REV H www.stryker.com...
NOTE: XXX - indicates language choice (tri = English/French/Spanish; bil = English/French) Item Part No. Part Name Qty. QDF27-0909-XXX Right Foot Siderail Cover Fascia QDF27-0910-XXX Left Foot Siderail Cover Fascia QP27-1831 LBS Lens Foot Siderail VV23A9C12HL Pan Head Tapping Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Right Foot Siderail Fascia with iBed QDF27-2031-XXX Left Foot Siderail Fascia with iBed QDF27-1834 LBS Cable Foot Siderail QDF2080 Inline Splice QDF27-1562 iBED Electronic Board Lens QP27-1831 LBS Foot Siderail Lens VV23A9C12HL Pan Head Tapping Screw Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Headboard and Footboard Assembly L27-028 Rev E (Reference Only) Item Part No. Part Name Qty. 27-1547 Footboard Assembly QP27-1077 Head Panel Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Cable Tie VE30A1G Nylon Hex Locknut VW10A06 Flat Washer 0056-028-000 Black TPR Bumper VV33A1G32 Pan Head Machine Screw 27-2770 Spring Load Connector VV33A1E16 Pan Head Machine Screw QE71-1367-T Label, Footboard Serial Number Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Footboard Assembly without iBed POSITION THE SCREEN TO HAVE QDF75-0290 BOARD THE YELLOW CABLES UPWARD ASSEMBLY CONNECTION OF THE TOUCHSCREEN ON J3 BOARD BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Footboard Assembly with iBed POSITION THE SCREEN TO HAVE THE YELLOW CABLES UPWARD QDF75-0290 BOARD ASSEMBLY CONNECTION OF THE TOUCHSCREEN ON J3 BOARD BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE Return To Table of Contents www.stryker.com 2141-209-002 REV H...
QDF2053 Red Electric Tape 75-0664 InTouch 4.0 Touchscreen Software QE71-1382-E InTouch 4.0 Software Label VV23A9C16HL Pan Head Tapping Screw 27-2840 LCD Screen 27-2698 Protection Tape 27-2841 Shim QDF2147 Double Face Tape Return To Table of Contents 2141-209-002 REV H www.stryker.com...
18 T5 POSITION THE SCREEN TO HAVE THE YELLOW CABLES UPWARD CONNECTION OF THE TOUCHSCREEN ON J3 BOARD QDF75-0290 BOARD ASSEMBLY BOARD ORIENTED TO HAVE THE WRITTING ON THE RIGHT SIDE VISIBLE Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Foot Control Frame 27-2698 Protection Tape VV23A9C12HL Pan Head Tapping Screw 5212-300-802 Label, FCC VV23A9C20HL Pan Head Tapping Screw 27-2841 Shim VVB3A9N24PF Bosscrew Screw VV23A9C16HL Pan Head Tapping Screw QDF2147 Double Face Tape Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Bed Extender Pad with Position Pro Mattress - DM64196-XXX 35” 4” 6” Bed Extender Pad with XPRT™ Mattress - DM64197-XXX 35” 5-1/2” 6” Return To Table of Contents www.stryker.com 2141-209-002 REV H...
IV Pole Ass’y, Dual Head End, Permanent - FA64202-XXX L64-107-XXX Rev A (Reference Only) HEAD OF BED TOQUE 75±10 Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Monitor Tray Assembly - FA64214-XXX Item Part No. Part Name Qty. QDF64-1300 Tray QDF64-1301 Transport Strap QDF64-1302 Fastener Sleeve QDF64-1303T Tray Support Pole Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Part No. Part Name Qty. 64-0647 Security Chain QDF64-1059 Bottle Holder QE71-1350-XXX Manufacturer Sticker QE71-0601 Maximum Load Sticker VR11H46 Dome Head Pop Rivet 64-1285 Oxygen Bottle Holder Socket VG10B0628 Spring Pin Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Pendant Assembly with Motion Control - FA64228-XXX L64-146-XXX Rev A (Reference Only) Motion Control Item Part No. Part Name Qty. QDF64-1240 Removable Pendant with Motion Control QE14399-T Manufacturer Sticker Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Pendant Assembly with Motion Control/NC - FA64226 L64-118 Rev 2 (Reference Only) Nurse Call Motion Control Item Part No. Part Name Qty. QDF64-1238 Removable Pendant with Nurse Call and Motion Control QE14399-T Manufacturer Sticker Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Pendant Assembly with Motion Control/Smart TV - FA64227 L64-119 Rev 3 (Reference Only) Smart TV Motion Control Item Part No. Part Name Qty. QDF64-1239 Removable Pendant with Motion Control and Smart TV QE14399-T Manufacturer Sticker Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Motion Control and Smart TV QE14399-T Manufacturer Sticker Pendant Clip Assembly - FA64186-XXX L64-111-XXX Rev A (Reference Only) Item Part No. Part Name Qty. QE71-1350-XXX Manufacturer Sticker QDF64-1282 Additional Removable Grip Return To Table of Contents 2141-209-002 REV H www.stryker.com...
OL270050 Rev 04 (Reference Only) Item Part No. Part Name Qty. 27-2046P Left Bumper Support 27-2056P Right Bumper Support QPC27-2059 Wheel, 1 3/4” Diameter VE30A1N Nylon Locknut VW10C122802 Nylon Washer 27-2235Z Bumper Spacer VB15A1N44 Bolt Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Traction Sleeve Assembly, 4 in. x 1/2 in. - FA64215-XXX L64-134-XXX Rev A (Reference Only) Item Part No. Part Name Qty. 64-1262C Traction Sleeve 4” x 1/2” QE71-1350-XXX Manufacturer Sticker VB15A1N32-S Hex Bolt VW10A08 Flat Washer Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Traction Sleeve Assembly, 4 in. x 3/4 in. - FA64216-XXX L64-135-XXX Rev A (Reference Only) Item Part No. Part Name Qty. 64-1263C Traction Sleeve 4” x 3/4” QE71-1350-XXX Manufacturer Sticker VB15A1N32-S Hex Bolt VW10A08 Flat Washer Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Traction Sleeve Assembly, 8 in. x 1/2 in. - FA64217-XXX L64-136-XXX Rev A (Reference Only) Item Part No. Part Name Qty. 64-1264C Traction Sleeve 8” x 1/2” QE71-1350-XXX Manufacturer Sticker VB15A1N32-S Hex Bolt VW10A08 Flat Washer Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Traction Sleeve Assembly, 8 in. x 3/4 in. - FA64218-XXX L64-137-XXX Rev A (Reference Only) Item Part No. Part Name Qty. 64-1265C Traction Sleeve 8” x 3/4” QE71-1350-XXX Manufacturer Sticker VB15A1N32-S Hex Bolt VW10A08 Flat Washer Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Traction Sleeve Assembly, 6-1/2 in. x 3/4 in. - FA64219-XXX L64-138-XXX Rev A (Reference Only) Item Part No. Part Name Qty. 64-1266C Traction Sleeve 6-1/2” x 3/4” QE71-1350-XXX Manufacturer Sticker VB15A1N32-S Hex Bolt VW10A08 Flat Washer Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Tube Support OL270224-XXX Rev 01 (Reference Only) Item Part No. Part Name Qty. 27-2518P Tube Support QE71-0498 Triangular Sticker QE71-1248-XXX Sticker to Drainage Tube VV83A9G24 Pan Phillips Head Screw Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part number: 27-2287/27-2288 (Reference Only) (27-2287 shown) Item Recycling/Material Code Important Information (QDF27-1097) Brake Control Board (QDF27-1099) Siderail Nurse Control Board Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: OL270311 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1429) Nurse Call Board (QDF75-0270) Room Interface Board Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part number: OL270302 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0270) Room Interface Board (27-2661) Left IR Module (27-2662) Right IR Module Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: OL270032 (Reference Only) Item Recycling/Material Code Important Information (QDF2060) Smart TV Board Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part number: OL270263-XXX (Reference Only) Item Recycling/Material Code Important Information (QDF27-1099) Siderail Nurse Control Board Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: 27-2547-XXX (Reference Only) Item Recycling/Material Code Important Information (27-2548) Zoom ® Control Board Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part number: OL270327-XXX Item Recycling/Material Code Important Information (QDF75-0010) Main Menu Board (QDF27-1097) Brake Control Board (QDF75-0290) Touch Board Assembly (27-2480) LCD Display Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: OL270326-XXX (Reference Only) Item Recycling/Material Code Important Information (QDF75-0010) Main Menu Board (QDF75-0630) USB Wi-Fi Board (QDF27-1097) Brake Control Board (QDF75-0290) Touch Board Assembly (27-2480) LCD Display Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part Number: 27-2688 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0440) DC Power Control Board (QDF9188) 12V 18Ah Battery Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: 27-2546 (Reference Only) Item Recycling/Material Code Important Information (QDF27-1430) Zoom® Board (QDF75-0240) Zoom® Interface Board Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Recycling Passports Assembly Part number: 27-2661 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0310) IR Board Return To Table of Contents www.stryker.com 2141-209-002 REV H...
Recycling Passports Assembly Part number: 27-2662 (Reference Only) Item Recycling/Material Code Important Information (QDF75-0310) IR Board Return To Table of Contents 2141-209-002 REV H www.stryker.com...
Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by Stryker, products or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product materially and...
Page 252
Manufactured For Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 2006/XX XXXX-XXX-XXX REV X www.stryker.com 2018/03 2141-209-002 REV H www.stryker.com...