Faults

If the fault cannot be rectified according to the faults table, Hettich customer services must be informed.
Please state the type of centrifuge and the factory serial number. Both values are visible on the centrifuge type plate.
MAINS RESET:
- Mains switch OFF for longer than 10 secs.
- Mains switch ON.
Fault
Display
No display
---
System reset
- 2 -
Balance error
- 3 -
Kommunication
- 4 -
Overload
- 5 -
Overvoltage
- 6 -
Undervoltage
- 8 -
Excess temperature
- 9 -
Controller watchdog
- C -
Lid error
- d -
Short circuit
- E -
Configuration
- F -
Cause of fault
No voltage
Mains input fuses defective.
Failure of power supply during a run
Unbraked runout.
Balance error on the motor axis due to
weight differences in the rotor loading.
Fault in control unit or power unit.
Unbraked runout.
Motor or motor control defective.
Supply voltage outside tolerance (see
Technical Data)
Unbraked runout.
Excess temperature switch in motor has
triggered.
Unbraked runout.
Fault in control unit
Unbraked runout.
Unbraked rundown,
after standstill lid release
Short circuit in control unit / power unit.
Incorrect machine version
Remedy
Check distribution voltage.
-
Check mains input fuses.
-
Mains switch ON.
-
When stationary, open lid and
-
press
key.
START
Open lid when rotor is stationary.
-
Eliminate balance error.
-
Perform mains reset when
-
stationary.
Perform mains reset when
-
stationary.
Perform mains reset when
-
stationary.
Check supply voltage.
-
When rotor is stationary, open lid
-
using emergency unlocking (see
Emergency Unlocking chapter).
Allow motor to cool down.
-
Perform mains reset when
-
stationary.
Perform mains reset after
-
standstill.
Perform mains reset when
-
stationary.
− Inform Customer Service.