Rotor removal / installation

Removal of haematocrit rotor:
1.
Open the lid and disconnect the centrifuge from the power supply.
2.
Remove the rotor lid by pressing the press lock.
3.
Unscrew the fixing nut with the supplied spanner.
4.
Pull the haematocrit rotor off the motor shaft.
4.
Installation of the haematocrit rotor:
1.
Insert the haematocrit rotor vertically onto the motor shaft. The motor shaft dog has to fit in the rotor slot.
2.
Tighten the fixing nut with the supplied spanner.
The fixing nut must always be firmly tightened to the motor shaft.