Laser Imager Installation - GE CT ProSpeed Series Installation Manual

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 15
3–6 LASER IMAGER INSTALLATION
Do not use the Pixel Clock signal when connecting a scanner to a camera.
Install a laser imager as follows (Refer to Illustration 3–20):
1.
Connect the ground cable to the rear OC frame and screw it securely into place, M4 Allen screw.
2.
Connect the co–axial video signal cable to the SV–AMP board, CN4, from the laser imager.
It may be necessary to use a HumEliminator depending on the type/model of laser imager.
3.
Connect the host control cable, if host control is used, to the MISC2 board, CN5.
4.
Connect the power cable, for FUJI laser cameras only, to one of the open terminals at TM1, No.9 and 10.
See Illustration 3–20, lower.
5.
Isolate any automatic film developer, if used, from the main body of the laser imager.
CT PROSPEED SERIES INSTALLATION
NOTICE
3-26
2124011