GE CT ProSpeed Series Installation Manual page 74

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 15
3–7–2 OC Modification (continued)
4.
Power cable Connection in OC
a. Four power cables are provided, one Black long cable, one Black short cable, one White long cable and one
White short cable.
b. Connect the four power cables as shown in Illustration 3–26.
5.
Signal Cable connection
a. Connect the OC internal cables by referring to Illustration 3–24. (There are 12 cables.)
b. Unscrew the DC III rear cover's 4 screws and remove the rear cover.
c. Connect the signal cable (for DC III control) and the video signal cable to the DC III. See Illustration 3–26 and
3–27.
The video signal cable goes to the Video Amp of the DC III.
The signal cable for DC III control goes to CN2 of DCCM2 ASSY of the DC III.
This signal cable has a ground line. Connect this ground line to the screw which is fixed to the OC nest. Twist
the ground line for the DC III side around the signal cable, then cover it with vinyl tape for isolation. See Illus-
tration 3–25.
TO
OCCM4
CN5
FIX THE GROUND TERMINAL WITH SCREW
FROM THE MAIN NEST OF THE OC.
CT PROSPEED SERIES INSTALLATION
FIX THE GROUND TERMINAL
WITH VINYL TAPE FOR ISOLATION.
SIGNAL CABLE FOR DC III
ILLUSTRATION 3–25
3-36
2124011
DC III
TWISTED
TO
DCCM2
CN2