GE CT ProSpeed Series Installation Manual page 71

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 15
3–7–2 OC Modification (continued)
b. OVL2 ASSY (Slot # 12)
Verify that the Resistor modules at location 17A1, 17A2, 19A1 and 19A2 are installed correctly.
17A1
17A2
19A1
19A2
c. IFB2 ASSY (Slot # 13)
This board has no jumper and no switches to adjustment.
d. FCP2A ASSY (Slot # 15)
Set all the Dip switches (SW 14A1) to OFF.
e. FCP1 ASSY (Slot # 16)
This board has no jumper and no switches to adjustment.
f. FCP3B ASSY (Slot # 17)
Set the Dip switches (SW 14A1) as follows.
STANDARD
FMD OPTION (DC III : Slot #18)
g. FMD ASSY (Slot # 18)
Set all the SW 08A1 Dip Switches to OFF.
D SW08A1 – FMD Address Mapping. Normally set as illustrated below
ON
1
2 3 4
330 W (U1222 RL)
. . . .
330 W (U1233 RL)
. . .
1 K W (U1310 RL)
. . . .
'1'
'0'
B0 : Not used. Set to '0'
B1 : Not used. Set to '0'.
B2–B3 : FMD Address Mapping on SYSTEM BUS
B3(SW1)
0
0
1
1
CT PROSPEED SERIES INSTALLATION
1
2
3
OFF
OFF
OFF
OFF
OFF
ON
B2(SW2)
Address(Hex)
0
1800000 – 2FFFFFF
1
3000000 – 47FFFFF
0
2000000 – 37FFFFF
1
3800000 – 4FFFFFF
3-33
2124011
4
OFF
OFF

Hide quick links: