3–1–1 Setting Coupling Terminal for Input Voltage (continued)
Input Voltage 200 V can not be used for systems having a 350 mA/120 kV output, or in the US.
There is a reserved metal plate at the right side of each TRANSFORMER COUPLING
TERMINAL. The plate is for 200v only.
When set up is for 200 V, it is necessary to adjust the metal plates. Height of these plates is adjusted
by using metal washers found at the right side of each TRANSFORMER COUPLING TERMINAL.
See Illustration 3–10.