3–1–1 Setting Coupling Terminal for Input Voltage (continued)
6.
Remove the rear cover, after removing the fan connector (brown connector) and the ground cable.
See Illustration 3–1 (C).
a. Verify that the rear cover is installed in step, and that the fans' connector is properly connected before applying
power to PDU.
7.
Remove the right side cover, after removing the fan connector (brown connector) and the ground cable.
See Illustration 3–1 (D).
A. REMOVE FOUR(4) HOIST
RINGS.
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
: FRAME GROUND POINT
FAN CONNECTOR
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
B. REMOVE THE TOP
COVER.
GROUND CABLE
D. REMOVE THE RIGHT
SIDE COVER.
POWER DISTRIBUTION UNIT FRONT VIEW