3–1–1 Setting Coupling Terminal for Input Voltage (continued)
8.
Set the voltage in the Power Distribution Unit (PDU) as follows:
a. Check the supplied voltage at Power Distribution Box.
Eight (8) levels of input RATING voltage (480, 460, 440, 415, 400, 380, 208, 200V) can be handled.
Input Voltage 208 V or 200 V can not be used for systems having a 350 mA/120 kV output.
For the System installed in the U.S.A., only 480V can be used as a PDU input voltage.
b. Remove the plastic cover from the transformer coupling plate ; See Illustration 3–2 (E).
c. Set the PDU input RATING voltage according to the appropriate setting shown in Illustration 3–3  Illustra-
tion 3–10.
The voltage is set by metal tap plates on the transformer coupling plate. See Illustration 3–2 (F).
E. REMOVE THE PLASTIC COVER
All connection types are delta. Therefore, there is no N(neutral) pin.
Plain Washer
   Â
NOTICE
NOTICE
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
Å Å Å Å Å Å
POWER DISTRIBUTION UNIT REAR VIEW
ILLUSTRATION 3–2
Note
Nut
Star Washer
Metal Tap Plate
3-5
CT PROSPEED SERIES INSTALLATION
F. TRANSFORMER COUPLING PLATE
Notice!:
The star washer MUST be
installed between nut and plain
washer to prevent loosening.
If the nut is loosened, a large
contact resistance might cause
the attaching parts to burn out.
2124011