3–5–1 Installation (continued)
HDD
5.
Change the CUSTOM PARAMETER for the MOD as follows. Refer to SYSTEM FUNCTIONAL CHECK Manual,
Section 5 (or 6), Load From Cold Procedure.
a. Using the BOOT Floppy, enter the Maintenance menu, and select CUSTOM PARAMETER program.
b. Use Change Parameters to set MOD to Available.
Do not set Optical Disk to available unless the site also has the 12" Optical Disk system installed.
c. After exiting CUSTOM PARAMETER program, reboot the system.
6.
Refer to the Operator Manual, Image Management section, and perform the MOD System Initialize, Archive,
Retrieve and cartridge handling operations.
7.
Refer to the Operators Manual, Image Management section to determine the required settings for Function
Menu. Pay special attention to Auto Store Device, OD Special Store, Auto Delete, Min Vacant Image Space for
Scan, OD Packing, and OD system No. Set these options as the site requires (default settings usually suffice).
The small screwdriver included in the MOD installation kit is used to provide a means of ejecting a
cartridge from the drive with power turned off. Insert the screwdriver in the Manual Eject hole (to the
left of the Busy LED), and turn clockwise about 10 turns. Be sure that power is OFF before performing
Manual eject.
+12VDC
PS
OC FRONT VIEW
5VDC ADJUSTING TRIMMER
ILLUSTRATION 3–19
Note
Note
3-24
CT PROSPEED SERIES INSTALLATION
H
V. ADJ.
ADJUSTER
2124011