GE CT ProSpeed Series Installation Manual page 148

Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV11
5 2 1
Gantry Anchoring
1.
Affix a piece of tape to the 5 mm concrete drill bit (Illustration 5-1) in order to assure 5 cm depth of anchor nut
hole.
2.
Make a hole at the anchor position shown in Illustration 5-2, using electric drill with the bit prepared in the
previous step, so that the lower edge of the tape just touch the floor surface.
ANCHOR
(Bolt during anchoring only)
ADJUSTER
BUBBLE LEVEL POSITION
5cm
5mm CONCRETE DRILL BIT PREPARATION
ILLUSTRATION 5-1
ADJUSTER
GANTRY ANCHOR POSITION
ILLUSTRATION 5-2
5-4
CT PROSPEED SERIES INSTALLATION
TAPE
GANTRY
2124011
TO TABLE