4220-Pgu And 4225-Pmu Output Limitations; Configure The Pgu, Pmu, And Rpm Using Tests - Keithley 4200A-SCS User Manual

Pulse card
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Figure 68: PMU Waveform Capture (measurement overflow on the 10 mA current range)

4220-PGU and 4225-PMU output limitations

In addition to the maximum output of the PMU and PGU instrument cards (see
determines pulse voltage across DUT
the number of large amplitude pulse transitions within a period of time. The 4200A-SCS system
enforces limits on the quantity and amplitude of waveforms that the 4220-PGU and 4225-PMU may
generate. A test exceeding these internal limits generates error code -830 (see pulse_init in
Model 4200A-SCS LPT Library Programming for more information). To fix this problem, increase
pulse period, decrease voltage amplitude, or both.

Configure the PGU, PMU, and RPM using tests

To configure and control the PGU, a user test module (UTM) is needed. You can also use UTMs to
configure and control the PMU and RPM preamplifier. For UTM programming, see "LPT commands
for PGUs and PMUs" in Model 4200A-SCS LPT Library Programming.
You can use interactive test modules (ITMs) to configure and control a PMU and RPM preamplifier.
Refer to "Configure a simple test" in Model 4200A-SCS Clarius User's Manual for information on
configuring an ITM for the PMU.
4200A-PMU-900-01 Rev. A December 2020
Section 3: Setting up PMUs and PGUs in Clarius
(on page 4-15)), the pulse instrument cards also have a limit for
DUT resistance
3-39