Section 2: Connections

PMU common connections

Common LO for the PMU is the outer shells of the two SMA connectors. With an SMA cable
connected (see following figure), common LO is the outside shield of the cable.
Because pulsing requires high frequency signal propagation, reduce cable inductance by minimizing
the loop area of the connection to the device under test (DUT). See the following figure and the
following figure for a graphic definition of the loop area for the coaxial cabling.
Do not use the GNDU as common low for the PMU to avoid creating a large loop area. When using
the GNDU, an inductive loop area is created when the HI and LO leads are separated. Fast rise times
(dt), high current (di), and large inductances (L) can cause voltage overshoots, oscillations, and
ringing in the high-speed measurement circuit. This is based on Lenz's law: V = L di/dt.

Shield connections

For multiple PMU channels, you should connect the shields (common LO) from all PMU channels as
close as possible to the DUT. You reduce inductance by minimizing the loop area of the shield
connections. The figure in
the
Local sensing
supplied cabling.
2-2
Figure 2: PMU common low terminals
Using an adapter cable to connect pulse card to DUT
(on page 2-15) figures illustrate proper shield connection schemes using the
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
4200A-PMU-900-01 Rev. A December 2020
(on page 2-14) and