Section 3: Setting up PMUs and PGUs in Clarius
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Figure 59: Two-channel test with three points (PMU1-1 and PMU1-2 displayed)

Higher channel count test example

This example shows how the Pulse Timing Preview is useful for higher channel count tests. The
following figure shows a four-channel test, reflecting the operation mode and voltages given in the
following table. When the channel count is higher, using the zoom feature provides a more detailed
view of the individual channel.
Expanded View zoom
On the Expanded View graph, drag the cursor to define the area to magnify (the cursor changes to a
magnifying glass). Note that each channel has a unique color and line width. When the channels
overlap, narrower lines are shown on top of the wider lines. To return to normal magnification,
double-click the graph or select Refresh. Multiple levels of zoom are supported, with a double-click
returning each level. Note that the Refresh option restores normal magnification (shows the Entire
Test waveform).