Use The Rpm To Switch The Smu, Cvu, And Pmu; Introduction; Equipment Required - Keithley 4200A-SCS User Manual

Pulse card
Hide thumbs Also See for 4200A-SCS:
Table of Contents

Advertisement

In this section:

Introduction .............................................................................. 6-1

Equipment required .................................................................. 6-1

Update the RPM configuration in KCon ................................... 6-2
Device connections .................................................................. 6-3
Set up the measurements in Clarius ........................................ 6-5
Introduction
The 4225-RPM Remote Amplifier/Switch Module is an accessory for the 4225-PMU UltraFast Pulse
Measure Unit that allows you:
To extend the current measurement ranges of the PMU to the 100 nA range.
To switch between the instruments without changing cables.
This section provides an example of how to use the 4225-RPM to switch the SMU, CVU, and
4225-PMU and make dc I-V, C-V, and pulsed I-V measurements to a single device without having to
reconnect the device between measurements.
For this example, you:
Make connections from two SMUs, one CVU, and the two-channel PMU to the inputs of two
4225-RPMs.
Make connections from the outputs of the two 4225-RPMs to a diode.
Generate dc I-V, C-V, and pulsed I-V measurements.
Equipment required
One 4200A-SCS with the following instruments:
Two 4200-SMUs, 4201-SMUs, 4210-SMUs, or 4211-SMUs
Two 4200-PAs
One 4210-CVU or 4215-CVU
One 4225-PMU
Two 4225-RPMs
Four 4200-TRX-2 or 4200-MTRX-2 triaxial cables (supplied with SMU)
Four CA-447A SMA cables (supplied with CVU)

Use the RPM to switch the SMU, CVU, and PMU

Section 6