4. From the Terminal Settings pane and Advanced Terminal Settings dialog box, you can adjust the
voltage source and current measurement parameters, as needed.
Figure 30: Terminal Settings pane and Terminal Settings Advanced dialog box

PMU operation modes

The operation mode determines what type of test is run on the terminal. Selecting the appropriate
mode simplifies configuration options.
A Pulse Sweep and Pulse Step function are identical. However, in a test where two or more PMUs
are used, the sweep for one PMU is performed on each step of the pulse step of the other PMU. A
pulse step requires that at least one channel be configured for Pulse Sweep.
A Pulse Train outputs one or more pulses of the same magnitude and base level.
The DC Gnd operation mode is used for the PMU GND terminal.
When there are two or more PMUs in the test system, the
used to select one of them as the master PMU.
3-4
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Sweep Master
4200A-PMU-900-01 Rev. A December 2020
(on page 3-23) option is