Section 3: Setting up PMUs and PGUs in Clarius

Low Range

Available when the Measure Range is set to Limited Auto. This sets the minimum range that the
instrument uses.

Current Sample Waveform

Makes current measurements on the waveform. Available when the Test Mode is set to Waveform
Capture.

Voltage Spot Mean High

Makes voltage measurements on the amplitude. Available when the Test Mode is set to Pulse IV.
The following figure shows an example of a spot mean measurement on pulse high (amplitude) and
pulse low (base level). For a more detailed example, refer to

Voltage Spot Mean Low

Makes voltage measurements on the base level. Available when the Test Mode is set to Pulse IV.
The following figure shows an example of a spot mean measurement on pulse high (amplitude) and
pulse low (base level). For a more detailed example, refer to

Voltage Sample Waveform

Makes voltage measurements on the waveform. Available when the Test Mode is set to Waveform
Capture.

Report Timestamps

Available when the Test Mode is set to Waveform Capture. When Report Timestamps is enabled,
each measurement includes a timestamp and the waveform is graphed (voltage/current versus time)
in the Analyze graph.
3-10
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Figure 37: Spot mean measurements
Figure 38: Spot mean measurements
Test Mode (PMU)
(on page 3-16).
Test Mode (PMU)
(on page 3-16).
4200A-PMU-900-01 Rev. A December 2020