Section 6: Use the RPM to switch the SMU, CVU, and PMU
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Figure 138: Test Settings pane and the Test Settings Advanced dialog box for the Pulse-Diode
test
11. Select OK to accept the changes.

Run the test

Once the tests have been configured, you can execute every test under the device.
To run the tests for the device:
1. Select the diode device in the project tree.
2. Verify that the check boxes for the tests and the device are selected.
3. Select Run. The three tests run sequentially and the RPM automatically switches the outputs
between the SMU, CVU, and PMU.
4. Select the Analyze pane to see the results.
As the instruments switch between tests, the LEDs at the top of the RPMs change color. When the
output is connected to the SMU, the LED is blue. When the output is connected to the CVU, the LED
is red. When the output is connected to the PMU, the LED is green, which is also the default state.