Section 6: Use the RPM to switch the SMU, CVU, and PMU
Two CA-547-2A RPM interconnect cables (supplied with RPM)
Four CA-534-24B triaxial cables
Two 237-TRX-T triaxial tees
One shielded test fixture with connection to GNDU (Force LO)

Update the RPM configuration in KCon

The KCon application is used to manage the configuration of the 4200A-SCS, including the
4225-RPM. Before using an RPM for automatic switching, you must update the RPM configuration in
KCon. This associates the instruments connected to each RPM and enables automatic switching
between tests.
To update the RPM configuration in KCon:
1. Make sure your device under test is disconnected from the RPM output terminals.
2. Close the Clarius application.
3. Open the KCon application.
4. Select Update.
5. Select Save.
6. Close KCon.
7. Open Clarius.
6-2
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Figure 119: Update the RPM configuration in KCon
4200A-PMU-900-01 Rev. A December 2020