Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
To enable connection compensation:
1. Select the test.
2. Select Configure.
3. Select the terminal to be compensated.
4. In the right pane, select Terminal Settings.
5. Select Advanced. The PMU Advanced Terminal Settings dialog box is displayed.
6. Select either Short Connection or Offset Current Correction. Refer to the following figure.
7. Select OK.
8. To disable connection compensation, clear either Short Connection or Offset Current
Correction, then select OK. When disabled, connection compensation values are not applied to
DUT measurements.
Load-line effect compensation (LLEC) for the PMU
Load-line effect compensation (LLEC) is only performed for standard pulse I-V testing using PMU
measure ranges. It is not performed when using 4225-RPM measure ranges (10 µA). The active
RPM circuitry provides its own analog LLEC (assuming there is a short cable from the RPM to the
DUT).
4200A-PMU-900-01 Rev. A December 2020
Figure 21: Enabling connection compensation
Section 2: Connections
2-21