Section 3: Setting up PMUs and PGUs in Clarius
F
Voltage measure overflow and current
measure overflow
G
Current measure range
H
Channel number and voltage
measure range

Compensation Short Connection

Use to enable or disable short connection compensation. Refer to
page 2-18) for detail on setting up and using connection compensation.

Load Line Effect Compensation

Use to enable or disable load-line effect compensation (LLEC). Refer to
compensation (LLEC) for the PMU
compensation.

DUT Resistance (R DUT)

The DUT resistance value. Refer to
4-15) for detail on calculating DUT resistance.
3-14
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
0 = No overflow
1 = Negative voltage overflow
2 = Positive voltage overflow
4 = Negative current overflow
5 = Negative voltage and negative current overflow
6 = Positive voltage and negative current overflow
8 = Positive current overflow
9 = Negative voltage and positive current overflow
A = Positive voltage and positive current overflow
0 = 100 nA (RPM only)
1 = 1 µA (RPM only)
2 = 10 µA (RPM only)
3 = 100 µA
4 = 1 mA (RPM only)
5 = 10 mA
6 = 200 mA
7 = 800 mA
1 = Ch1, 10 V
2 = Ch2, 10 V
5 = Ch1, 40 V
6 = Ch2, 40 V
(on page 2-21) for detail on setting up and using load-line effect
DUT resistance determines pulse voltage across DUT
PMU connection compensation
Load-line effect
4200A-PMU-900-01 Rev. A December 2020