15 S
PECIFICATIONS
OLP-85

SPECIFICATIONS

15
S
PECIFICATIONS
76

OLP-85 specifications

Adjustable wavelength range
Calibrated wavelenghts
Photo diode
Fiber types
Display range
Resolution
Maximum input power level
1)
Intrinsic uncertainty
Overall measurement
2) 3)
uncertainty
4)
Wavelength detection
4)
Modulation detection
1) Valid for calibrated wavelengths 1310 nm, 1490 nm, 1550 nm at reference con-
ditions, at -22 dBm (CW), 23 °C ±3 K, 9 µm test fiber with SC/PC ceramic con-
nector. Intrinsic uncertainty for calibrated wavelengths 850 nm (-20 dBm),
980 nm (-22 dBm) and 1625 nm (-22 dBm) ±0.25 dB (±6 %)
2) -75 dBm to +26 dBm, -5 °C to +45 °C
3) Overall measurement uncertainty for 800 to 1700 nm:
800 nm to 1300 nm: ±0.7 dB ±25 nW
1300 nm to 1550 nm: ±0.4 dB ±5 nW
> 1550 nm: ±0.7 dB ±5 nW
4) With Viavi optical light source, 800 nm to 1625 nm: level > -50 dBm
850 nm, 980 nm, 1310 nm,
1490 nm, 1550 nm, 1625 nm
-75 dBm to +26 dBm
850 / 980 nm:
1310 / 1490 / 1550 nm:
1625 nm:
Automatic switching and displaying up to
4 wavelengths simultaneously
270 Hz, 1 kHz, 2 kHz
800 nm to 1700 nm
filtered InGaAs
9/125 to 62.5/125
0.01 dB, 0.001 µW
+30 dBm
±0.15 dB (±3.5 %)
±0.35 dB ±25 nW
±0.25 dB ±5 nW
±0.35 dB ±5 nW