Technical Data; Sitrans Fus080 - Siemens SITRANS FUS080 Operating Instructions Manual

Ultrasonic flowmeters
Table of Contents

Advertisement

Technical data

10.1

SITRANS FUS080

Table 10- 1
Description
Enclosure design/material
Enclosure
Wall mounting kit
Cable glands
Ambient temperature
Storage temperature
Transducer / sensor cables
Mechanical vibration
Power supply
Display
Push button
Measuring function
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
Technical data, FUS080
Specification
Fiber glass reinforced polyamide in light-gray color
IP67 according to EN 60529 and DIN 40050 (NEMA 4X/6)
Terminal box made of fiber glass reinforced polyamide in light-gray
color with 2 + 2 cable glands (output, supply, and transducer
cables) and a stainless steel bracket for wall or pipe mounting
In wall mounting kit:
-5 °C ...+60 °C (23 °F...140 °F)
−40 °C...+85 °C (−40 °F...+185 °F)
2 g, 1...800 Hz sinusoidal in all directions according to IEC 68-2-6
Battery:
Mains:
One push button for toggling between display information
or
4 x M20 polyamide cable glands (2 for mains and output
cables, 2 for the transducer coaxial cables)
Coaxial cable, impedance 75 Ohm, ∅ app. 6mm;
Cable length: Max. 30 meter between transmitter and
transducer (factory sets 15 or 30 m (49.21 or 98.43 ft)
Replaceable 3.6 V LiSOCl (Lithium Thionyl Chloride)
Double battery pack 32 Ah or single back-up battery 16 Ah;
Battery change interval: 6 years at 60 °C (140 °F) operation
(depending on used functions)
87 ... 265 V AC (50 ... 60 Hz)
LCD, 8 digits, additional 2 digits and symbols for status
information
Units:
Volume unit: m
(default display unit)
3
Flow unit: m
/h (default display unit)
3
Alarm codes: F 1 to F 9
0.5 Hz battery mode
15 Hz mains-powered