Siemens SITRANS FUS080 Operating Instructions Manual page 37

Ultrasonic flowmeters
Table of Contents

Advertisement

Read all parameters
Before any parameterization is done it is necessary to read all parameters from the device
into the offline table of SIMATIC PDM. The offline table merely contains default data.
1. Open PDM device driver.
2. Select "Upload to PC/PG .." Select "Execute even if the device TAG does not match the
3. Store factory set default values on your local PC ("File" → "Export") in order to be able to
Read, write device data
Only parameters (data) shown with white background can be changed.
Note
Status field
 "Changed" indicates off-line data not yet stored in the device.
 "Loaded" indicates actual device data.
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
project data TAG.", and click "OK" to read all parameters to the offline table.
After closing the dialog box, all loaded parameters should show "Loaded" in the status of
the PDM table. For a complete parameter list, see "Parameter list" (Page 69).
retrieve default settings.
Commissioning
6.4 Commissioning via PDM
35