Siemens SITRANS FUS080 Operating Instructions Manual page 39

Ultrasonic flowmeters
Table of Contents

Advertisement

Set sensor-related parameters
1. Select "Device"→"Pipe geometric assistant" in order to calculate:
2. Define sensor-specific parameters based on data from the sensor measurement report
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
– Rn for path 1 and 2 (pipe constant in line 304 and 305)
– Calibration Factor (Flow-Profile-Factor for retrofit, line 310)
– Diameter (line 306)
and the selection made at ordering:
– Select metric (m, m/s) or imperial (in, ft/s) units for data entry. Internal calculation in
the device is made in metric units only.
– Enter outer pipe diameter (D
– Flow velocity (V) is automatically calculated using entered pipe data and given Q
(PDM parameter row 10).
– Enter media viscosity in cSt (e.g. 1 cSt for water at 20 °C).
– Enter the pipe roughness value.
The pipe-roughness default value is automatically set according to pipe material or
can be entered manually by selecting "User defined" (range: 0.0000 to 0.0100 m).
Pipe material
Smooth plastic pipe (select PVC)
Polished stainless steel pipe
Standard carbon steel pipe (select STEEL)
Rusty carbon steel pipe
Concrete pipe (Select CONCRETE)
) and average pipe wall thickness (t).
u
Typical roughness
0.0001 m
0.0001 m
0.0004 m
0.001 ... 0.002 m
0.002 ... 0.005 m
Commissioning