Perm Oven
The Perm Oven screen provides status and alarms for the permeation oven,
if installed. If an item being monitored goes outside the lower or higher
alarm limit, an alarm is activated.
Use the  and  buttons to move up and down and the  and 
buttons to move left and right.
Home Screen>Settings>Health Check>Status and Alarms>Perm Oven
The Perm Oven screen contains the following information:
Across:
Down:
Perm Oven: This column lists items associated with the perm oven.
Value: Displays the current value for each item.
Low Alarm: Displays low alarm for each item.
High Alarm: Displays high alarm for each item.
Units: Displays units for each item.
Perm Oven Temperature: Displays the current perm oven
temperature. User can adjust low and high alarm limits.
Board Communication: Displays OK/Fail for board communication
status.
Power Supply: Displays OK/Fail of power supplies. Power supply
goes red if any voltages are outside their limits. No voltage rows
ever get highlighted.
2.5 V Diagnostic: Displays current voltage readings. Alarm limits
are not changeable.