Thermo Scientific 42iQ Instruction Manual page 113

No-no2-nox analyzer

Advertisement

Thermo Scientific
Down:
Manual Entry: Shows current value. If the input board cal cell is
pressed, user can manually change the value.
Auto Cal: When pressed, the auto-calibration process is initiated.
Please allow up to 5 minutes for calibration to complete.
End Cal: When pressed, the auto calibration is interrupted and the
value does not change.
PMT Voltage: Displays PMT voltage.
Frequency: Displays frequency.
Zero Offset Freq: Displays zero offsets frequency.
Default Values: When pressed, the default values are saved.
Home Screen>Settings>Measurement Settings>Advanced Measurement
Settings>Reaction Chamber Settings>Continue to PMT Calibration
Screen
The PMT Calibration screen contains the following information:
Across:
Description: Defines the actions the user can do.
PMT Supply: User can manually adjust the PMT supply value,
adjust target concentration and read calibration concentration,
start/stop automatic calibration, read voltage, frequency and zero
offset frequency values, and reset default values.
Current PMT: Displays the current PMT voltage value.
Units: Displays units for each item.
Operation
Settings
42iQ Instruction Manual
3-87