Thermo Scientific 42iQ Instruction Manual page 205

No-no2-nox analyzer

Advertisement

Thermo Scientific
Problem
Alarm – Module
Temperature
Alarm – 5V/24V Step Board
Alarm – Analog I/O
Alarm – Digital I/O
Alarm – NO Bkg Check
Offset
Alarm – NO Span Check
Offset
Alarm – Ambient
Thermistor
Alarm – Chamber Pressure
Alarm – Bench
Temperature Short
Alarm – Bench
Temperature Open
Alarm – Cooler Current too
Low
Possible Cause
Action
Cable connection
Check that DMC cable is connected
properly. Reseat if needed.
Other alarm
Make sure the instrument
temperature is not too high or in
alarm.
Defective DMC PCB
Replace DMC board.
Cable connection
Check the cable connections to that
Step POL board.
Defective PCB
Replace Analog board.
Defective PCB
Replace Digital board.
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument
Recalibrate the instrument.
background
calibration failed
Incorrect high alarm
Verify the high limit is correct via
limit
Settings>Status and
Alarms>Concentrations screen.
Instrument span
Recalibrate the instrument.
calibration failed
Defective
Replace defective DMC board as
Flow/Pressure board
needed.
Defective
Replace defective DMC board as
Flow/Pressure board
needed.
Cable connection
Check the cable connection from the
heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the heater.
Defective heater
Replace heater as needed.
Defective DMC board
Replace defective DMC board as
needed.
Cable connection
Check the cable connection from the
heater to the DMC board.
Broken wire
Verify the wires are properly
connected to both sides of the heater.
Defective heater
Replace heater as needed.
Defective DMC board
Replace defective DMC board as
needed.
Defective cooler
Replace cooler as needed.
Troubleshooting
Troubleshooting Guide
42iQ Instruction Manual
6-5