Calibrate No X Low; Calibrate No 2 Low - Thermo Scientific 42iQ Instruction Manual

No-no2-nox analyzer

Advertisement

Calibrate NO
Calibrate NO
Thermo Scientific
3. When the responses stabilize, from the Home screen, choose
Calibration>Calibrate Span Coefficients>Calibrate Low Range NO
Span Coefficient.
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Low Range NO Span
Coefficient button will show what the span coefficient will be set to if
the Calibrate button is pressed.
4. Enter the NO calibration gas concentration using the pushbuttons, and
then press Calibrate to calculate and save the new low range NO span
coefficient based on the entered span coefficient.
Low
Use the following procedure to calibrate the NO
x
calibration gas.
1. Press the Back button to return to the Calibration menu and choose
Calibrate Low Range NO
The user sets the span concentration by pressing the Edit Span
Concentration button. The Calculated Low Range NO
Coefficient button will show what the span coefficient will be set to if
the Calibrate button is pressed.
2. Verify that the low NO
low NO calibration gas concentration plus any known NO
3. Enter the NO
and then press Calibrate to calculate and save the new low range NO
span coefficient based on the entered span coefficient.
Low
Use the following procedure to calibrate the NO
2
calibration gas.
1. Adjust the O
produce a decrease in the low NO concentration equivalent to about
80% of the URL of the low NO
90% of the low NO concentration determined in the
Low" procedure.
2. Press the Back button to return to the Calibration menu and choose
Calibrate Low Range NO
Span Coefficient.
x
calibration gas concentration is the same as the
x
calibration gas concentration using the pushbuttons,
x
generator in the GPT system to generate sufficient O
3
range. The decrease must not exceed
2
Span Coefficient.
2
Calibration in Dual Range and Auto Range Mode
channel to the NO
x
Span
x
2
channel to the NO
2
"Calibrate NOx
42iQ Instruction Manual
Calibration
x
impurity.
x
2
to
3
4-19