    Apply the single point measurement method, and take measurements   
    continuously along the designated route, the intervals should be less 
    than 5mm, and take the minimum value as the workpiece's thickness. 
9.2 Pipeline measurement method 
During the measurement, make the probe's crosstalk interlayer plate be 
perpendicular or parallel to the axial line of the pipeline. For a pipeline 
with larger diameter, the probe's crosstalk interlayer plate should be 
perpendicular to the axial line of the pipeline, but for pipeline with small 
diameter, User should perform measurements making the crosstalk 
being both parallel and perpendicular to the axial line of the pipeline and 
take the minimum readout as the thickness value.   
10. Maintenance and precautions   
10.1 Power check 
When the power is low, the low battery indicator will appear, at this 
moment User should replace the battery in time, or it will affect the 
measuring accuracy. The backlight cannot be switched on for a long 
time, because it is a big consumer of electricity. 
Note: if the unit did not used for a long time, please take out of the 
battery to avoid leakage to damage the unit. 
10.2 Precautions 
10.2.1 General precautions 
The unit should avoid strong vibration, do not let it in an excessively 
humid environment, plug in or out the probe should hold the jacket to 
avoid the core wire of the probe damaged.   
10.2.2 Precaution during the measuring   
1). During the measurement, only the measuring icon appears and   
    displayed stable, it can be regarded as a good measurement. 
2). If there are large quantity coupling agents attached on the 
      measured surface, when taking away the probe, it will cause error,   
      so when the measurement is completed, please move the probe 
      away from the measured surface quickly. 
3). If the probe wears out, it will cause the displayed value unstable,   
      please replace the probe. 
 
 
 
 
 
 