Gain
The gain can be adjusted over a wide range. The setting of the gain is crucial
in order to obtain valid reading during the measurement process. Too much
gain may result in erroneous measurements, by detecting on noise rather
than the actual material back wall itself. Too little gain may result in
detection on an undesirable section of the waveform.
The gain will also be represented in both the B‐SCAN and DIGITS views.
The following steps outline how to enable and adjust the GAIN.
1) Press
once to illuminate GAIN.
2) Use keys
(+1),
adjustment)to adjust value.
Delay
The starting DELAY is the value displayed on the bottom lower left side of
the display in both RF+ and HALF+ views. It is the minimum thickness value
that can be viewed on the display.
The following steps outline how to enable and adjust the DELAY.
This is DELAY
indicator.
1) Press
once to illuminate DELAY.
2) Use keys
(+1),
adjustment)to adjust value.
(‐1),
(‐coarse adjustment ),
(‐1),
(‐coarse adjustment ),