Blank
BLANK is used in T‐E mode. Make change of Blank to conceal those useless
echoes that influence measuring result. When making changes, the red line
at the left bottom conner changes accordingly. User could take it as for
reference and achieve the best measuring effect.
The following steps outline how to enable and adjust the BLANK.
This red line is
BLANK.
1) Press
once to illuminate BLANK.
2) Use keys
(+1),
adjustment)to adjust value.
Saving the setting
Once all the parameters are set, there is two ways to save the setting.
1) The user can Press
When turn on the gauge next time, this setting will be displayed.
2) Or the user can Press
SETTING for further use while measuring the same workpiece.
8.2.7 Rectification
DC‐5000 supplys two display views of A‐Scan. RF+ mode shows both the
positive and the negative peaks. HALF+ mode shows the upper of complete
waveform only.
The following steps outline how to select the RECTIFICATION.
1) Press the
key to illuminate SET。
(‐1),
(‐coarse adjustment ),
to store this setting into the current location.
to store this setting into the A‐SCAN CUSTOM
‐ 24 ‐
(+ coarse