Default; Measuring Technology; Measuring Methods - DGC DC-5000 Instruction Manual

Ut thickness gauge
Table of Contents

Advertisement

8.4.4 Default   
During the usage, when User can not ensure why the problems comes out 
and with some questions on setting, he can use this function to make the 
parameters to restore the factory status to eliminate any abnormal because 
of the parameters setting.   
 
8.4.5 Information   
The screen displays the Transducer Number.   
 
 
9. Measuring technology 
9.1 Measuring methods 
The unit provides many measuring methods. 
1). Single point measuring method:   
    Use the probe to measure any point Of the workpiece to be measured 
    and the displayed value is the thickness. 
2). Two point measuring method: 
    Perform two measurements on the same point of the measured   
    Surface. In the second measurement, splitting plane of the probe 
    should be 90 degree, take the minimum as the thickness value. 
3). Multiple point measurement method: 
    Perform several measurements In a circle about 30mm in diameter 
    and take the minimum value as the thickness value. 
4. Continuous measurement methods:   
 
 
‐ 30 ‐ 
 
 
 

Advertisement

Table of Contents
loading

Table of Contents