DC-3000 SERIES ULTRASONIC THICKNESS GAUGE
4.
Continuous measurement methods:
apply the
single
point
measurement method, and take measurements continuously along the
designated route, the intervals should be less than 5mm, and take the
minimum value as the workpiece's thickness.
【9.2】 Pipeline measurement method
During the measurement, make the probe's crosstalk interlayer plate be
perpendicular or parallel to the axial line of the pipeline. For a pipeline
with larger diameter, the probe's crosstalk interlayer plate should be
perpendicular to the axial line of the pipeline, but for pipeline with small
diameter, users should perform measurements making the crosstalk being
both parallel and perpendicular to the axial line of the pipeline and take
the minimum readout as the thickness value.

10. Maintenance and precautions

【10.1】 Power check
When the power is low, the low battery indicator will appear, at this
moment users should replace the battery in time, or it will affect the
measuring accuracy. The backlight cannot be switched on for a long time,
because it is a big consumer of electricity.
Notice: if the instrument did not used for a long time, please take out of
the battery to avoid leakage to damage the instrument.
【10.2】Precautions

10.2.1 General precautions

The instrument should avoid strong vibration, do not let it in an
excessively humid environment, plug in or out the probe should hold the
jacket to avoid the core wire of the probe damaged.

10.2.2 Precaution during the measuring

① During the measurement, only the measuring icon appears and
displayed stable, it can be regarded as a good measurement.