7. Quick startup guide
Step one:
Selecting the Transducer and Probe zero & calibration
Right now, there is probe S1025 available and its information is as below:
Freq.
Type
(MHz)
S1025
1
Plug in transducer; turn on the main unit using the
auto calibration of the transducer, thus eliminating the need for an on‐block
zero. After turning on the gauge, the screen flashes the Series No. and
software version, and then, it comes into the measurement mode directly.
If users turning on the unit without the transducer, screen will remind to
"Plug in the probe". At this moment, please insert a transducer into the
socket, the gauge directly comes into the measurement mode after Auto
calibration.
Notice: Please use the standard transducer offered. Otherwise the unit does
not work normally and displaying "Error".
If users feel the thickness reading is incorrect during the measurement,
please put the probe in the air, and press
time.
Notice: For Auto calibration, make sure the transducer is not coupled to the
test piece when the gauge is first turned on and there is no couplant on the
end of the transducer. The transducer should also be at the room
temperature, clean without any noticeable wear.
Step Two:
Treatment of the measured surface
When the surface to be measured is too rough or rusty heavily, please
perform the treatment according to the following methods:
1. Clean the measured surface by grinding, polishing or filing, etc. or use
coupling agent with high viscosity for that.
2. Use coupling agents on the workpiece surface to be measured.
3. Take multiple measurements around the same testing point.
Dia.
Meas. Rang
(mm)
(mm)
25
20~590
Working
Temp.
Used in the measurement of
<60℃
composite materials
for zero calibration at any
‐ 8 ‐
Application
Key. The gauge does