8.2.6 Manual measurement 
This feature allows user to make fine adjustment of RANGE, GAIN, DELAY, 
GATE, BLANK, and E‐BLANK manually in A‐SCAN mode. They can be set 
through toggle the HOT MENUS what is displayed at the lower left side of 
the screen. 
And once the above mentioned parameters are set, it will remain the same 
for B‐SCAN and DIGITS display.   
Range 
The range refers to the overall viewable range being displayed on the 
screen. Make change the range of display, shorten or enlarge it, finally 
serve User conveniently. 
The following steps outline how to enable and adjust the RANGE. 
1) Press
  once to illuminate RANGE. 
2) Use keys 
    adjustment)to adjust value.   
Gate 
GATE is used in both T‐E and E‐E modes. The purpose of GATE is to force 
DC‐5000 to measure the useful echo wave.   
The following steps outline how to enable and adjust the GATE. 
This red line is 
GATE 
1) Press
  once to illuminate GATE. 
2) Use keys 
    adjustment)to adjust value.   
 
(+1), 
(‐1), 
(+1), 
(‐1), 
 
  (‐coarse adjustment ),
 
  (‐coarse adjustment ),
‐ 22 ‐ 
  (+ coarse 
  (+ coarse 