8.2.3 Probe Calibration 
It causes error during the primary stage of usage or operate for long time. 
User should make probe calibration during following three aspects are 
happened. 
1. The probe itself or the temperature variation, 
2. System error caused by the match between the unit and the transducer, 
3. Calculation error caused by the sound velocity set in the unit is different 
    from that of the actual material. 
This feature requires a sample piece of the specific material to be measured, 
the exact thickness of which is known. E.g. from having been measured by 
some other means. Or to use the build‐in standard test block (4MM) comes 
with DC‐5000. 
The following steps outline how to enable and set up this feature: 
1) Apply a drop of couplant on the transducer and place the transducer in 
    steady contact with the sample or standard test block.   
2) Having achieved a stable reading, remove the transducer.   
3) Press the
    Scroll through the sub menu items until Probe calibration is highlighted. 
    Press the
   
4) Press the 
    until the thickness value displayed matches the thickness of the test 
    block or sample piece. Press the 
5) Press the 
    verify the result.   
 
 
 
  key to illuminate SET, Use
  key to confirm.   
 
  or 
  key to adjust the displayed reading up or down,   
  key to exit setting, Test the block or sample piece again to 
 
 
 
  key to confirm. 
‐ 20 ‐ 
 