mA Output

When a SuSix® is connected to a power supply for the first time, the
mA output is automatically set to provide 4 mA at zero and 20 mA at the
theoretical maximum value.
Changes in the mA setting will not affect the relay output settings.
Note:
Both values can be set in the range 10 % to 120 %, making it
possible to increase or decrease the mA signal.
Units
Units
NTU (Turbidity)
FTU (Turbidity)
FNU (Turbidity)
EBC (Turbidity)
%
(Suspended solids)
mg/l (Suspended solids)
g/l
(Suspended solids)
Kg/l (Suspended solids)
Back
Select
The mA output is an active output, and the maximum load is 800 Ohm.
The upper limit for the mA output is 20,5 mA
3,75 mA indicates the mA output is not in use
3,5 mA indicates values under 4 mA level
20,5 mA indicates value higher than 20 mA level.