Turbidity and Suspended Solids Transmitter
3. Dip the SuSix
®
the fluid in which the suspended solids measurement is going to take
place.
Ensure sure that the extracted sample does not sediment (deposit) by
stirring continuously.
Important!
If the signal from the suspended solids sensor changes very
slowly on the display, make sure that the signal attenuation is not
set too high. Check likewise in the "Averaging" menu that the
response time is not set too high.
The recommended response time is 5 seconds.
EN 5.10 SuSix Manual 2007
sensor in the sample stirring continuously, or leave it in
25