Turbidity calibration
A complete, normalized linearization curve of the probe is installed by the
manufacturer. No further calibration is possible (or required).
Suspended solids calibration
There are no calibration standards pre-defined for the system. Each meas-
urement must be calibrated individually, and it is possible to define two
suspended solids curves for use in different liquids.
Suspended solids calibration requires a sample to be evaluated in a
laboratory in order to set the relation between the measured point and the
actual value of the sample.

Averaging

The "Averaging" option sets the time period within which the measure-
ments are smoothed and averaged.
The current measured value is compared with the standard deviation of
the previous measured values. This feature ensures that a sudden devia-
tion in the liquid composition is eliminated in the measurement
Clear "Adjust samples"
Clear "Adjustment sample"
No
Yes
Back
Select
1. Select "Yes" in the "Quality improvement" window.
EN 5.10 SuSix Manual 2007
Turbidity and Suspended Solids Transmitter
Quality improvement
No
Yes
Back
(from 1 sek to 120 min)
Back
Ok
Averaging
Select
>
Yes
Averaging
0001 sec
Select
>
60
accordingly
Units
Units
NTU (Turbidity)
FTU (Turbidity)
FNU (Turbidity)
EBC (Turbidity)
%
(Suspended solids)
mg/l (Suspended solids)
g/l
(Suspended solids)
Kg/l (Suspended solids)
Back
Select